Installation
Micro-Ion
®
Module Instruction Manual - 356007-GP
21
In
stallation
O
per
a
tion
Maintenance
Before You
Begin
Relay Wiring
The module has two trip point relays. The contacts are silver alloy-gold
clad, rated for 1 A at 30 Vdc. The relays can handle resistive or
non-inductive loads.
• Figure 2-6 illustrates the 9-pin trip point connector.
• Table 2-3 lists pin connections for the 9-pin male D connector.
Figure 2-6
Trip Point Relay 9-pin Connector
Table 2-3
Trip Point Relay 9-pin Connector Pins
Pin #
Function
Input/Output
Voltage Level
These voltage levels apply ONLY for resistive or non-inductive loads.
Pin 1
Relay 2: Normally open
Output
0 to 30 Vdc, 1 A
Pin 2
Relay 2: Common
Output
0 to 30 Vdc, 1 A
Pin 3
Relay 2: Normally closed
Output
0 to 30 Vdc, 1 A
Pin 4
Not used
No connection
Pin 5
Relay 1: Common
Output
0 to 30 Vdc, 1 A
Pin 6
Relay 1: Normally closed
Output
0 to 30 Vdc, 1 A
Pin 7
Relay 1: Normally open
Output
0 to 30 Vdc, 1 A
Pin 8
Not used
No connection
Pin 9
Not used
No connection
Summary of Contents for 356 Micro-Ion Plus Series
Page 2: ......
Page 24: ...24 Micro Ion Module Instruction Manual 356007 GP...
Page 58: ...Chapter 3 58 Micro Ion Module Instruction Manual 356007 GP...
Page 74: ...Index 74 Micro Ion Module Instruction Manual 356007 GP...
Page 75: ......