14
Series 275 Convectron Module with DeviceNet
Instruction Manual - 275502
2 Introduction/Specifications
2.2
Specifications
Table 2-1
Specifications
Function
Description
Measuring range for Air or N
2
(See Chapter 4 of this instruction manual
for correction curves for other gases.)
• 1 x 10
-4
to 1000 Torr
Operation in the range of 1x10
–4
to 1x10
-3
Torr will require periodic
verification of the zero by reducing system pressure to below 1 x 10
–5
Torr
• 1 x 10
–4
to 1300 mBar
• 1 x 10
–2
to 130 KP
a
Resolution
1 x 10
-4
Torr; 1 x 10
–4
mBar; 1 x 10
–2
pascal
Mounting position
Horizontal preferred
IP Rating
•
IP20
Power required
•
+11.5 Vdc to 26.5 Vdc
+11.5 Vdc, 0.22 A
+26.5 Vdc, 0.1 A
•
3.5 watts maximum
•
Protected against reversals, transients or over-voltages
Digital interface
DeviceNet
Baud rate
125K, 250K, or 500K, switch selectable
Setpoint
•
Single pole, double throw relay, silver alloy - gold clad contacts
•
UL rating: 1 A at 30 Vdc, 0.3A @ 12.5 VAC
•
Potentiometer adjustment
Connectors
5-pin Micro
9-pin D male
Ambient operating temperature range
0 °C to +40 °C (non-condensing)
Non-operating temperature range
-40 °C to +70 °C
Bakeout temperature range (non-operating)
85 °C max.
Package
•
Aluminum extrusion design with aluminum end plates
•
4.5” long x 1.7” wide x 2.5” high plus gauge port and fitting
Weight
340gm (12 oz) with 1/8 NPT fitting
Gauge tube internal volume
40 cc (2.5 cu in.)
Sensor wire
Platinum or Gold-plated Tungsten
Summary of Contents for 275420-1-GQ-T-00011
Page 2: ......
Page 26: ...4 Operation 26 Series 275 Convectron Module with DeviceNet Instruction Manual 275502 Notes...
Page 31: ...5 DeviceNet Operation Series 275 Convectron Module with DeviceNet Instruction Manual 275502 31...
Page 32: ...5 DeviceNet Operation 32 Series 275 Convectron Module with DeviceNet Instruction Manual 275502...
Page 50: ......
Page 51: ......