![MKS 2480HS Installation & Use Manual Download Page 8](http://html1.mh-extra.com/html/mks/2480hs/2480hs_installation-and-use-manual_1813478008.webp)
MKS MultiGas High-Speed Gas Sampling Cart
Additional
Safety
Information
for
the
MultiGas
Cart
CAUTION! LIFTING HAZARD! Two people are required to perform this
procedure due to size, position, and/or weight of a component.
This symbol indicates that a component is considered a “Heavy Object” and
requires a two-person lift. Please take caution in lifting the device and have
appropriate assistance when doing so. In addition, the NIOSH Lifting Formula
should be used in a Manual Handling Worksheet to evaluate the final installation
location of a component or equipment for the in-use location determined by the
end-user. This should be used to determine if the component could exceed the two-
person lift limitations and require a lift assist in order to position or remove it. Refer
to center of gravity drawings.
WARNING: BURN HAZARD! Allow system components to cool as directed
prior to performing maintenance to avoid personal injury.
Caution: Toxic vent must be connected prior to use. Using a 7/8 line, connect the toxic vent of
the sample cart to facility vent or vented outside of your facility.
Caution: Toxic Pressure relief vent line must be connected prior to cart use. Using a 1/2 in
line, connect the pressure relief line to facility vent or vented outside of your facility.
NOTE: Toxic Vent and Toxic Pressure relief should not be plumbed together. They must
remain dedicated lines up to the facility exhaust.
Installation and Use Manual 2011 APRIL 2 Rev. 1.5
8