MKS MultiGas High-Speed Gas Sampling Cart
Installation
Safety
Notice
Warning
All users of the equipment must read and understand all safety
information before installing or working on the MultiGas High-Speed
Gas Sampling Cart.
Facility
Requirements
Refer to the system dimensional drawings and center –of–gravity drawings provided with the
system.
Requirement
Specification
Recommended Installation/
Use Space
Minimum System Dimensions
Height: 71”
Width: 26”
Depth: 41”
Recommended Installation Space to allow for Connections, Service, and Operation:
Height: 77”
Width: 74”
Depth: 113”
System Power
115VAC, two (2) 15A circuits
The system requires two 115 VAC grounded outlets for power connections.
The outlets must supply a minimum of 10 Amps each.
Use of appropriate power extensions cords are required for the power load and the installation
work environment.
Nitrogen: 1 slpm, 0-20 psig, high purity
Gas Requirements
Calibration Span Gas: 1 slpm, 0-20 psig
Exhaust Requirements
A toxic vent and toxic pressure relief is required for handling the system exhaust gases as
necessary to prevent damage to personnel and equipment. Follow all local, national, and
applicable gas handling and exhaust requirements and regulations.
The toxic vent should be connected to a 25 ft 7/8” diameter exhaust line to minimize back
pressure on the system. The connection is located on the lower rear panel
A ½ in Toxic Pressure Relief exhaust line should be used to vent the pressure release valve in
the event of system over pressure. The connection is located on the lower rear panel
Operation and Storage
Conditions
Operating Temperature: 18 - 30°C
Installation and Use Manual 2011 APRIL 2 Rev. 1.5