5
2
Safety precaution
CAUTION
Have the person who operates the operating
table to operate it in a position where the
emergency stop switch can be immediately
pressed, and the patient’s condition can
constantly observed.
180°
Back plate
Head plate
Operator
position
Patient’s
head
Other
WARNING
●
Prohibited
•
Do not disassemble and/or modify the operating table. Otherwise, malfunction may
occur.
●
Patient position
•
When using the tabletop or accessories to secure a patient’s body position, always
observe the patient’s condition.
Being in the same position for long periods of time causes neuroparalysis or
bedsores.
●
Other medical electrical equipment to be used together with the operating table
• When using a high-frequency surgical equipment and/or a cardiac defibrillator
etc. with the operating table, refer to their operator's manual provided by the
manufacturers and observe the precautions and usage. Improper precautions
and usage may cause the operator or the patient to get burned and/or devices to
malfunction.
•
When using medical electrical equipment etc. with the operating table, check
that the operating table does not malfunction before using it. Electromagnetic
interference may result in malfunctioning of the operating table.
●
Allowable load
•
Do not apply a load which exceeds the allowable load*. The operating table may not
function, which may result in failures.
* Lift: 250 kg (550 lbs) / Lateral tilt and Trendelenburg: 250 kg (550 lbs)
●
Preventive maintenance and inspections
•
Make sure to inspect and maintain the operating table before and after use. The
operating table may require replacement of the parts due to significant wear,
deterioration, and/or breakage depending on the length of service and frequency of
use.
•
For preventive maintenance and inspections, contact your distributor or Mizuho
directly.
●
Antistatic measure
• Do not use the operating table on floors and/or with accessories that do not
possess static electricity countermeasures. This may impede surgical operations.
Summary of Contents for MOT-VS500DK
Page 3: ...App 1 Electromagnetic Compatibility 52 App 2 Glossary 55 ...
Page 4: ......
Page 60: ...Revision Record New release Ver 1 2020 07 03 ...
Page 61: ......