47
6
Specification
Dimension
Tabletop
2133 mm (83.9 in) (L) x 500 mm (19.6 in) (W) : Note 5
Base
1033 mm (40.6 in) (L) x 483 mm (18.9 in) (W) : Note 6
Weight
360 kg (793 lbs)
Allowable load
Lift: 250 kg (550 lbs) / Trendelenburg, Lateral tilt: 250 kg (550 lbs)
Transitable height and width
Height : 10 mm / Width : 80 mm
Operating environment
Temperature
10 to 40°C: Note 9
Humidity
30 to 75%: Note 9
Atmospheric pressure
700 to 1060 hPa: Note 9
Others
Allowable altitude for use is 3000 m or lower.: Note 9
Transportation and storage
Temperature
-10 to 50°C: Note 7
Humidity
10 to 85% (without moisture condensation) : Note 7
Atmospheric pressure
700 to 1060 hPa: Note 7
Service life
Under the specified maintenance and proper storage, 10 years: Note 8
Note 1: from the center position (where the side frame and the sliding cylinder’s line are aligned)
Note 2: Optional
Note 3: When the battery power is used
Note 4: Consecutive pressing of the switch of the control unit.
Note 5: Excluding the side rail
Note 6: Rough dimension
Note 7: Company standard (in case that appropriate maintenance and inspection is done)
Note 8: Based on Mizuho's own validation data
Note 9: IEC 60601-1, Medical electrical equipment - Part1: General requirements for safety