Operating Instructions MG3 Digital
Integration in Automated Installations
76609409EN-BA-MG3D-V1_6.docx 67/82
2010-08-02
6
Integration in Automated Installations
Integrating measuring instrument MG3 Digital in an automated installation, take care to
have the MG3 Digital ready signal always co-evaluated by a higher-level stored-program
controller (SPC).
There must be no triggering as long as the ready signal is “low” (measuring interval).
During this measuring interval, data is calculated, compared, and transferred to the
display and further interfaces.
In the following chapter, flow charts illustrate the time sequence of the signals by a
number of examples.
6.1
Flow Chart Channel 1
or
Channel 2 alone
high
high
high
low
low
low
Dia1-MG3
Current Output
outputs
limit values
accept / reject
pre-warn
value
Ready Output
Current signal
current1
or current 2
measuring delay
ready for new measuring
new sequence
no measuring possible!
adjusted measuring time