8
2. Place and tighten the Femoral Rod screw to lock
into place.
3. Slide the Guide Block onto the Guide Frame, then
attach cover to Guide Block and fasten.
4. Lock the Guide Block screw into place.
5. Attach Arc Attachment to Guide Block when
necessary.
PACKAGING
• For instruments that are provided in sets, instrument
trays may be used for containment during sterilization.
• Package trays/instruments with a barrier wrap
material in accordance with local procedures, using
standardized wrapping techniques such as those
described in ANSI/AAMI ST79, latest revision.
STERILIZATION Trays
• Place DePuy Mitek instrument in the appropriate
location within the DePuy Mitek sterilization tray. If a
specific location is not identified for an instrument,
it may be placed in the general purpose area (pin
mat) ensuring that the sterilant has adequate access
to all surfaces including difficult to reach areas,
lumens, etc.
• Use a validated, properly maintained and calibrated
steam sterilizer.
• Effective steam sterilization can be achieved using
the following cycles:
Cycle Type
Temperature
Minimum
Exposure Time
Pre-vacuum
132°C
(270°F)
4 min
Pre-vacuum
(outside the U.S.)
134 – 137°C
3 min
Drying Time: 30 minutes