Passport
®
Quick Reference Guide
To view and print an alarm event on a patient on the Passport 8 or Passport 12
bedside monitor:
a. Select
Review >>
(quick key) or from
Main Menu
(quick key)
b. Select
Events
tab
c. Scroll and select event
d. Select
Details
to view the 8, 16, or 32-second event*
e. Select
•
Record
to print to the integrated recorder
•
to print to the DPM Central Station/independent printer
f. Select
Main Menu
(quick key) or
to return to the main display
* Alarm Recording length configurable
Note:
The waveforms printed are determined by the Recorder and Print
Setup menus on the Passport 8 or Passport 12 bedside monitor.
DPM Central Station Interface:
Event storage and printing length on the
Passport 8 or Passport 12 bedside monitor and DPM Central Station are
independent.
Panorama Central Station Interface:
Event storage and printing length on the
Passport 8 or Passport 12 bedside monitor and Panorama Central Station are
independent.
Full Disclosure
The Passport 8 or Passport 12 bedside monitors have the ability to store
configured waveforms and associated numeric information for up to 48 hours.
To adjust waveforms to be stored on a patient:
a. Select
Review>>
(quick key) or from
Main Menu
(quick key)
b. Select
Full Disclosure
tab
c. Select
Save Waves >>
d. A
in the waveform box indicates waveform is selected
e. Select Confirm to return to
Full Disclosure
tab
f. Select
Main Menu
(quick key) or
to return to the main display