Parameters >>
(
or
Main Menu
quick keys)
d. Verify the computation constant or adjust the computation constant in
Setup>>
e. Inject solution when
Inject Now
message is displayed (Auto measuring
mode is enabled in
Setup >>
)
Or
e. Select
Start
and inject solution within 4 seconds to perform a cardiac
output measurement when
Ready for new set of measurement
is
displayed (Manual measuring mode is enabled in
Setup>>
)
Note:
Cardiac index is calculated if height/weight has been entered in the
Patient Setup menu
f. Wait 1 minute between injections to allow baseline stabilization
g. Perform additional Cardiac Output measurements as needed
h. Select measurement to include/exclude in the Cardiac Output average (a
in the box indicates cardiac output measurement is selected)
i. Select
1.
Cancel
to stop the current measurement
2.
Record
to record the most recent cardiac output measurement and
numeric data
j. Select
Main Menu
(quick key) or
to return to the main display
Or
j. Select
Calc>>
to access the Hemodynamic Calculations menu
Note:
Blood temperature alarms are inactive when a cardiac output
measurement is initiated.
Initiate Pulmonary Artery Wedge Pressure (PAWP)
Pulmonary Artery Wedge Pressure (PAWP) is a pressure measurement derived
from a PA catheter when the PA distal balloon is inflated and the catheter
advances and occludes a distal pulmonary artery. PAWP pressure is a reflection
of the pressure in the left ventricle at end-diastole.
To initiate a PAWP measurement on the Passport 8 or Passport 12 bedside