P/N: H-046-005847-00 Rev 3.0
17
Passport
®
Quick Reference Guide
monitor:
a. Insert IBP Module as needed
b. Select
PAWP
menu by selecting PA invasive blood pressure tile or from
Parameters
menu (
or Main Menu
quick keys)
Or
a. Select
PAWP
(quick key)
b. Select
Setup
to adjust PA sweep speed, PA scale, or reference waveforms
(as needed)
c. Inflate balloon and watch PA waveform for wedge waveform
d. Select
Freeze
to freeze the waveform once a satisfactory wedge
waveform is detected
e. Select
Adjust
to adjust the PAWP reference line to the desired position
f. Select
Confirm
to enter the PAWP measurement
g. Select
1.
Next Measurement
to start a new measurement as needed and repeat
previous steps
2.
Exit
to return to normal screen
h. Select
Calc>>
to access the Hemodynamic Calculations menu
Note:
PA alarms are inactive when a PAWP measurement is initiated.
Note:
Follow manufacturer’s suggested procedures and hospital policy for
PAWP balloon inflation.
Initiate Hemodynamic Calculations
Hemodynamic calculations are a set of values that are used to determine the
hemodynamic status of the patient.
To initiate hemodynamic calculations on the Passport 8 or Passport 12 bedside
monitor:
a. Select
CALC>>
from the C.O. menu or PAWP menu or
HEMODYNAMIC
from
CALCULATIONS
menu (
or
Main Menu
quick keys)
b. HR, Art (Mean), PA (Mean), CVP, and average CO measurements are
Summary of Contents for Passport 12