P/N: H-046-005847-00 Rev 3.0
11
Passport
®
Quick Reference Guide
Adjust Alarm Responses
To adjust a patient’s alarm responses on the Passport 8 or Passport 12 bedside
monitor:
a. Select parameter’s waveform or numeric tile to open parameters Setup
menu
b. Select
Alarm Setup >>
c. Select
•
On/Off
– Enable/disable the alarm’s responses
•
Level
– to adjust the audio and visual alarm indicators. Level options
include High (Red), Medium (Yellow), and Low (Yellow)*
•
Record
– to activate an automatic recording when the alarm occurs
•
Print (DPM Central Station Interface)
prints patient’s alarm settings to
designated laser printer
•
Default
– to restore preset settings
d. Select
Main Menu
(quick key) or
to return to the main display
Or
a. Select
Alarm Setup >>
from
(quick key) or
Main Menu
(quick key)
b. Select
Parameters
tab to change parameters or
ST
tab to change ST
c. Scroll and select parameter
d. Select
•
On/Off
– to enable or disable the alarm’s limits or responses
•
Level
– to adjust the audio and visual alarm indicators. Level options
include High (Red), Medium (Yellow), and Low (Yellow)*
•
Record
– to activate an automatic recording when the alarm occurs
•
Print (DPM Central Station Interface)
prints patient’s alarm settings to
designated laser printer
•
Default
– to restore preset settings
e. Select
Main Menu
(quick key) or
to return to the main display
*Parameter dependent