P/N: H-046-005847-00 Rev 3.0
14
Passport
®
Quick Reference Guide
Station, WorkStation, and ViewStation for a
preconfigured amount of time
.
Selecting
Mute
on the Panorama Central Station
will not
silence an alarm on
the Passport 8 or Passport 12 bedside monitor.
SILENCE ALL ALARMS
To silence
all alarms
on the Passport 8 or Passport 12 bedside monitor for a
preconfigured amount of time, select
Alarm Pause
key on front panel of
bedside monitor.
DMP Central Station Interface:
Selecting
Alarm Pause
on the Passport 8
or Passport 12 bedside monitor will silence active alarms on the Panorama
Central Station, WorkStation, and ViewStation for a
preconfigured amount of
time
. Selecting
Alarm Pause
on the Panorama Central Station will silence
alarms on the Passport 8 or Passport 12 bedside monitor.
Panorama Central Station Interface:
Selecting
Alarm Pause
on the Passport
8 or Passport 12 bedside monitor will silence active alarms on the Panorama
Central Station, WorkStation, and ViewStation for a
preconfigured amount of
time.
Selecting
Mute
on the Panorama Central Station
will not
silence alarms
on the Passport 8 or Passport 12 bedside monitor.
TRANSFER*
The Passport 8 or Passport 12 bedside monitors have the ability to transfer
patient data through a storage medium (USB Flash Drive)
To transfer patient data from one Passport 8 or Passport 12 bedside monitor to
another Passport 8 or Passport 12 bedside monitor:
a. Insert USB Flash Drive into USB port on back of monitor
b. Select
Patient Setup >>
using
Patient Setup
(quick key) or
Main
Menu
(quick key)
c.
Select Transfer to Storage Medium*
d. Select
OK
e. Remove USB Flash Drive once
Transfer to storage medium successful
message is displayed
f. Insert USB Flash Drive into USB port on back of new Passport 8 or
Passport 12 bedside monitor
Summary of Contents for Passport 12
Page 1: ...Quick Reference Guide ...