A-7
A.3 Data Storage
Trends
Trends: 96 hours, at 1 min resolution
Minitrends: 1 hour, at 1 s resolution
Parameter alarms
100 alarms and related parameter waveforms. The waveform
recording length can be 8s, 16s or 32s.
Arrh. events
80 arrhythmia events and relate waveforms. The waveform
recording length can be 8s, 16s or 32s.
NIBP measurements
1000 sets
12-lead ECG analysis results 10 sets
Full-disclosure waveforms
48 hours at maximum. The specific storage time depends on
the waveforms stored and the number of stored waveforms.
A.4 Wireless Network
Standards IEEE
802.11b/g
Frequency range
2.412 to 2.462 GHz
A.5 Measurement Specifications
The adjustable range of alarm limits is the same with the measurement range of signals
unless otherwise specified.