16.2 Preparing to Measure CO
2
16.2.1 Using a Sidestream CO
2
Module
1. Attach the watertrap to the module and then connect the CO
2
components as shown
below.
2. By default, the sidestream CO
2
module is in standby mode. Access the [
CO
2
Setup
]
menu to set [
Operating Mode
] to [
Measure
]. The [
CO
2
Startup
] message appears on
the screen.
3. After start-up is finished, the CO
2
module needs time to warm up to reach the operating
temperature. The message [
CO
2
Sensor Warmup
] is displayed. If you perform CO
2
measurements during warm-up, the measurement accuracy may be compromised.
4. After warm-up is finished, you can perform CO
2
measurements.
NOTE
z
To extend the lifetime of the watertrap and module, disconnect the watertrap and
set the operating mode to standby mode when CO
2
monitoring is not required.
Watertrap
Watertrap fixer
Sampling line
Summary of Contents for iPM-9800
Page 1: ...iPM 9800 Patient Monitor Operator s Manual ...
Page 2: ......
Page 44: ...3 14 FOR YOUR NOTES ...
Page 50: ...4 6 FOR YOUR NOTES ...
Page 58: ...5 8 FOR YOUR NOTES ...
Page 102: ...8 24 FOR YOUR NOTES ...
Page 108: ...9 6 FOR YOUR NOTES ...
Page 118: ...11 8 FOR YOUR NOTES ...
Page 128: ...12 10 FOR YOUR NOTES ...
Page 140: ...14 10 FOR YOUR NOTES ...
Page 166: ...17 10 FOR YOUR NOTES ...
Page 178: ...19 10 FOR YOUR NOTES ...
Page 188: ...20 10 FOR YOUR NOTES ...
Page 200: ...22 6 FOR YOUR NOTES ...
Page 206: ...23 6 FOR YOUR NOTES ...
Page 212: ...24 6 FOR YOUR NOTES ...
Page 260: ...A 28 FOR YOUR NOTES ...
Page 266: ...B 6 FOR YOUR NOTES ...
Page 299: ......
Page 300: ...PN 9211 20 87439 7 0 ...