14-9
14.7 Zeroing the Transducer
To avoid inaccurate pressure readings, the monitor requires a valid zero. Zero the transducer
in accordance with your hospital policy (at least once per day).Zero whenever:
A new transducer or adapter cable is used.
You reconnect the transducer cable to the monitor.
The monitor restarts.
You doubt the readings.
1. Turn off the stopcock to the patient.
2. Vent the transducer to the atmospheric pressure by turning on the stopcock to the air.
3. In the setup menu for the pressure (e.g. Art), select [
Zero Art >>
]
→
[
Zero
]. During
zero calibration, the [
Zero
] button appears dimmed. It recovers after the zero
calibration is completed.
4. After the zero calibration is completed, close the stopcock to the air and open the
stopcock to the patient.
NOTE
z
Your hospital policy may recommend that the ICP transducer is zeroed less
frequently than other transducers.
3-way stopcock
Pressure transducer
Pressure adapter cable
Monitor
Summary of Contents for iPM-9800
Page 1: ...iPM 9800 Patient Monitor Operator s Manual ...
Page 2: ......
Page 44: ...3 14 FOR YOUR NOTES ...
Page 50: ...4 6 FOR YOUR NOTES ...
Page 58: ...5 8 FOR YOUR NOTES ...
Page 102: ...8 24 FOR YOUR NOTES ...
Page 108: ...9 6 FOR YOUR NOTES ...
Page 118: ...11 8 FOR YOUR NOTES ...
Page 128: ...12 10 FOR YOUR NOTES ...
Page 140: ...14 10 FOR YOUR NOTES ...
Page 166: ...17 10 FOR YOUR NOTES ...
Page 178: ...19 10 FOR YOUR NOTES ...
Page 188: ...20 10 FOR YOUR NOTES ...
Page 200: ...22 6 FOR YOUR NOTES ...
Page 206: ...23 6 FOR YOUR NOTES ...
Page 212: ...24 6 FOR YOUR NOTES ...
Page 260: ...A 28 FOR YOUR NOTES ...
Page 266: ...B 6 FOR YOUR NOTES ...
Page 299: ......
Page 300: ...PN 9211 20 87439 7 0 ...