A-14
Nellcor SpO
2
Module
Parameter
Specifications
Standards
Meet standards of ISO9919
Measurement range
1 to 100%
Accuracy
70 to 100%: ±2% (measured in adult/pediatric mode)
70 to 100%: ±3% (measured in neonate mode)
0% to 69%: Not specified.
Refreshing rate
1 s
SPO
2
averaging time
8 s, 16 s
*: When the SpO
2
sensor is applied for neonatal patients as indicated, the specified accuracy range
is increased by ±1%, to compensate for the theoretical effect on oximeter measurements of fetal
hemoglobin in neonatal blood.
A.5.4 PR
Alarm limit
Range (bpm)
Step (bpm)
PR High
(low limit +2) to 240
PR Low
25 to (high limit-2)
1
PR from Mindray SpO
2
Module
Measurement range
20 to 254 bpm
Resolution 1
bpm
Accuracy ±3
bpm
Refreshing rate
2 s
SPO2 averaging time
7 s (when sensitivity is set to High)
9 s (when sensitivity is set to Medium)
11 s (when sensitivity is set to Low)
PR from Masimo SpO
2
Module
Measurement range
25 to 240 bpm
Resolution 1
bpm
Accuracy
±3 bpm (measured without motion)
±5 bpm (measured with motion)
Refreshing rate
1 s
SPO2 averaging time
2-4 s, 4-6 s, 8 s, 10 s, 12 s, 14 s, 16 s
Summary of Contents for iPM-9800
Page 1: ...iPM 9800 Patient Monitor Operator s Manual ...
Page 2: ......
Page 44: ...3 14 FOR YOUR NOTES ...
Page 50: ...4 6 FOR YOUR NOTES ...
Page 58: ...5 8 FOR YOUR NOTES ...
Page 102: ...8 24 FOR YOUR NOTES ...
Page 108: ...9 6 FOR YOUR NOTES ...
Page 118: ...11 8 FOR YOUR NOTES ...
Page 128: ...12 10 FOR YOUR NOTES ...
Page 140: ...14 10 FOR YOUR NOTES ...
Page 166: ...17 10 FOR YOUR NOTES ...
Page 178: ...19 10 FOR YOUR NOTES ...
Page 188: ...20 10 FOR YOUR NOTES ...
Page 200: ...22 6 FOR YOUR NOTES ...
Page 206: ...23 6 FOR YOUR NOTES ...
Page 212: ...24 6 FOR YOUR NOTES ...
Page 260: ...A 28 FOR YOUR NOTES ...
Page 266: ...B 6 FOR YOUR NOTES ...
Page 299: ......
Page 300: ...PN 9211 20 87439 7 0 ...