Accuracy of Measurement B-1
Appendix B Accuracy of
Measurement
Within the selected field range, the measurement accuracy is ensured within the range mentioned
above.
The accuracy specifications are performance in the worst conditions, or based on the real
test for the system.
Depth and time
Item
Error range
Depth
≤±
4% of full scale
M-mode image
time
≤±
0.3% of full scale
Accuracy
of
Measurement
Item
Error range
Distance/Depth
≤±
4%, or <2 mm if measured value is less than 40 mm
Area (Trace)
≤±
8%, or <130
if measured value is less than 1600
2
mm
2
mm
Area (Ellipse)
≤±
8%, or <130
if measured value is less than 1600
2
mm
2
mm
Angle
≤±
3%
%stenosis
Max. 1.0
≤±
7%
Time/Motion
Measurements
Item
Error range
Distance
≤±
4%, or <2 mm if measured value is less than 40 mm
Time
≤±
1%
Heart Rate
≤±
5%
Slope
≤±
5%
Volume
Measurements
Item
Error range
Volume
≤±
12%, or <8000
if measured value is less than 64000
3
mm
3
mm