![Mindray DP-8300 Operator'S Manual Download Page 72](http://html1.mh-extra.com/html/mindray/dp-8300/dp-8300_operators-manual_1800872072.webp)
Item
Type
Description
TSI
Drop-down list
To set the default TSI.
B Map/THI Map
Button
To configure the B Map/THI Map. Refer to the
following segments for detailed operations.
Depth Drop-down
list
To
set the default depth.
THI Check
box
Checked/Unchecked: For transducers that support
harmonic, you can set harmonic as default
frequency when this item is checked.
Frequency
Drop-down list
To set the default frequency.
Scan Angle
Drop-down list
To set the default scan angle: 0, 1, 2, 3.
F.Position
Drop-down list
To set the default focus position.
Focus Number
Drop-down list
To set the default number of focuses.
Gain
Drop-down list
To set the default gain.
THI Gain
Drop-down list
To set the default THI gain. This item is available
for transducers that support harmonic only.
Line Density
Drop-down list
To set the default scanning density: Hi Frm Rate,
Hi Density 1 and Hi Density 2.
L/R Flip
Radio button
To set the default horizontal flipping mode: Left,
Right.
U/D Flip
Radio button
To set the default vertical flipping mode: Up, Down.
Rotation Drop-down
list
To set whether to rotate the image or not: 0, 90,
180, 270.
B IP
Drop-down list
To set the default B IP: 1~8.
THI IP
Drop-down list
To set the default THI IP: 1~8. This item is
available for transducers that support harmonic
only.
B IP/THI IP
Button
To configure the parameters of B IP/THI IP. Refer
to the following segments for detailed operations.
The THI IP is available for transducers that support
harmonic only.
M
Item
Type
Description
Speed
Drop-down list
To set the default M scanning speed.
M Map
Button
To configure the M Map. Refer to the following
segments for detailed operations.
M Gain
Drop-down list
To set the default M Gain.
M Map
Drop-down list
To set the default M Map.
Edge Enhance
Drop-down list
To set the default parameter of Edge Enhance.
M IP
Drop-down list
To set the default M IP.
M IP
Button
To configure the M IP. Refer to the following
segments for detailed operations.
8-8 Preset
Summary of Contents for DP-8300
Page 1: ...DP 8500 DP 8300 Digital Ultrasonic Diagnostic Imaging System Operator s Manual Basic Volume ...
Page 2: ......
Page 8: ......
Page 20: ......
Page 36: ......
Page 54: ......
Page 58: ......
Page 82: ......
Page 86: ......
Page 92: ......
Page 94: ......
Page 96: ......
Page 101: ...P N 046 000308 00 V1 1 ...