![Mindray DP-8300 Operator'S Manual Download Page 29](http://html1.mh-extra.com/html/mindray/dp-8300/dp-8300_operators-manual_1800872029.webp)
2
Preparing the System
This chapter introduces how to prepare and get familiar with the system.
2.1
2.1.1
Setup and Connections
Moving and Placing the System
Please read and understand the safety precautions before moving and placing the system.
1. Turn off the power and disconnect the peripheral devices.
2. Unlock the four casters.
3. Move the system to the desired position using the handle.
4. Leave at least 20cm clearance at the back and two sides of the machine.
5. Lock the four casters.
CAUTION:
Ensure enough clearance at the back and both side of the machine,
otherwise failure may happen because of the increasing temperature
inside the machine.
2.1.2 Connecting the Power Cord
Connecting the power cord
WARNING:
1. Do not connect this system to outlets with the same circuit
breakers and fuses that control current to devices such as life-
support systems. If this system malfunctions and generates an
overcurrent, or when there is an instantaneous current at power
ON, the circuit breakers and fuses of the building’s supply circuit
may be tripped.
2. Do not connect the three-wire power cable of the machine to a two-
wire plug without grounding protection phase, otherwise electric
shock may happen.
1. Ensure that the external power supply system meets the requirements and the power switch of
the system is turned off.
2. Attach the connector of the power cord to the system.
3. Insert the power plug securely into the wall outlet.
Connecting the chassis earth lead
The chassis earth lead is used to put the ultrasonic system to earth and make the ultrasonic
system equipotential with the other devices connected to the system.
1. Connect the connector of the chassis earth lead to the equipotential terminal of the
ultrasonic system.
2. Put the clip of the chassis earth lead to earth.
Preparing the System 2-1
Summary of Contents for DP-8300
Page 1: ...DP 8500 DP 8300 Digital Ultrasonic Diagnostic Imaging System Operator s Manual Basic Volume ...
Page 2: ......
Page 8: ......
Page 20: ......
Page 36: ......
Page 54: ......
Page 58: ......
Page 82: ......
Page 86: ......
Page 92: ......
Page 94: ......
Page 96: ......
Page 101: ...P N 046 000308 00 V1 1 ...