Gas Module 3 Operating Instructions

0070-10-0696-02

2 - 9

Operations

Gas Monitoring with Gas Module 3

GM: O

2

 Zero Error

Appears when the system has been 
unable to successfully zero the O

2

 

sensor.

Manually start zeroing the 
system again. If problem 
persists, contact Mindray DS 
Technical Support.

GM: N

2

O Zero Error

Appears when the system has been 
unable to successfully zero the N

2

sensor.

Manually start zeroing the 
system again. If problem 
persists, contact Mindray DS 
Technical Support.

GM: Agent Zero Error

Appears when the system has been 
unable to successfully zero the 
anesthetic agent sensor.

Manually start zeroing the 
system again. If problem 
persists, contact Mindray DS 
Technical Support.

GM: Pump Off

Appears when the system has turned 
off the pump due to a pneumatic 
error.

Restart the pump from the 

Gas Menu

. If problem 

persists, contact Mindray DS 
Technical Support.

GM: Agent Mismatch - 
HAL 

Appears when the system detects 
Halothane as the primary agent and 
the manually selected agent is not 
Halothane.

Match the Agent 
administered with the Agent 
selected, or select 

Agent 

Auto ID

.

GM: Agent Mismatch - 
ISO

Appears when the system detects 
Isoflurane as the primary agent and 
the manually selected agent is not 
Isoflurane.

Match the Agent 
administered with the Agent 
selected, or select 

Agent 

Auto ID

.

GM: Agent Mismatch - 
ENF

Appears when the system detects 
Enflurane as the primary agent and 
the manually selected agent is not 
Enflurane.

Match the Agent 
administered with the Agent 
selected, or select 

Agent 

Auto ID

GM: Agent Mismatch - 
SEV

Appears when the system detects 
Sevoflurane as the primary agent and 
the manually selected agent is not 
Sevoflurane.

Match the Agent 
administered with the Agent 
selected, or select 

Agent 

Auto ID

GM: Agent Mismatch - 
DES

Appears when the system detects 
Desflurane as the primary agent and 
the manually selected agent is not 
Desflurane.

Match the Agent 
administered with the Agent 
selected, or select 

Agent 

Auto ID

GM: Unknown Agent

Appears when the system detects a 
gas that does not match the 
spectroscopic signatures of the five 
known anesthetic agents

Use recognized agent

GM: Cannot Zero... 
RETRYING

Appears when the bedside monitor 
requests 

Zeroing

 (either on the 

automatic cycle or by a user request) 
and the Gas Module is unable to 
initialize the cycle

Allow system to retry without 
intervention. If problem 
persist, contact Mindray DS 
Technical Support.

GM: CO

2

 Uncalibrated

Appears after an unsuccessful 
calibration attempt of the CO

2

 sensor. 

The numeric data for CO

2

 will appear 

as - - -, and the CO

2

 waveform will be 

a flatline

Ensure proper gas mixture is 
attached tightly and regulator 
is on. Repeat calibration 
procedure. If problem 
persists, contact Mindray DS 
Technical Support.

MESSAGE/PROBLEM

REASON

SOLUTION

Summary of Contents for Accutorr 3

Page 1: ...Operating Instructions Gas Module To Purchase Visit Avobus com or call 1 800 674 3655...

Page 2: ...registered trademark of Mindray DS USA Inc Spectrum is a U S registered trademark of Mindray DS USA Inc Spectrum OR is a U S trademark of Mindray DS USA Inc Copyright Mindray DS USA Inc 2008 All righ...

Page 3: ...perations 2 1 Gas Monitoring with Gas Module 3 2 1 Pre use Test 2 3 Gas Monitor Calibration Passport 2 Spectrum Spectrum OR 2 4 Passport 2 Spectrum 2 4 Spectrum OR 2 6 Gas Module Troubleshooting 2 8 U...

Page 4: ...Gas Module 3 Operating Instructions 0070 10 0696 02 1 ii This page intentionally left blank To Purchase Visit Avobus com or call 1 800 674 3655...

Page 5: ...configuration of the bedside monitor Patents This device is covered under one or more of the following U S Patents 6 589 028 6 896 713 and foreign equivalents Possession or purchase of this device do...

Page 6: ...e Gas Module 3 on the floor Connect only a bedside monitor to the same MPSO as the Gas Module 3 Do not overload the MPSO WARNING Do not connect other equipment to the same MPSO with the Gas Module 3 a...

Page 7: ...damaged during operation there is a risk that bacteria and or mucus may contaminate the Gas Module WARNING Do not use Adult Pediatric type water traps and or sampling lines with neonates to avoid high...

Page 8: ...ca gel have been included CAUTION Contamination with CO2 N2O or Anesthetic Agent in the air surrounding the Gas Module 3 may cause significant measurement errors Notes NOTE Potential hazards due to er...

Page 9: ...rent AC Direct Current DC Data Input Output Interference may occur in the vicinity of equipment marked with this symbol Crossed out wheelie bin indicates separate treatment from general waste at end o...

Page 10: ...Gas Module 3 Operating Instructions 0070 10 0696 02 vii This page intentionally left blank To Purchase Visit Avobus com or call 1 800 674 3655...

Page 11: ...ir Adult Pediatric P N 0202 00 0182 10 Neonate P N 0202 00 0181 10 The Water Trap Assembly is used to capture moisture drawn in with the patient sample The Water Trap Reservoir must be emptied and rin...

Page 12: ...ower Cord 6 Exhaust Port This panel mount coupling is used for attaching a gas scavenging system P N 0997 00 0923 or P N 0997 00 0984 to the Gas Module 3 7 External Interface Port A communication inte...

Page 13: ...e monitors allows for the measurement of anesthetic gases O2 N2O and CO2 levels Measurement can be acquired via a nasal cannula non intubated for oxygen and CO2 only or through a sampling line connect...

Page 14: ...ient For intubated patients connect the sample line to the breathing circuit Refer to instruction provided in the sample line packets 3 Connect the other end of the nasal cannula or sample line to the...

Page 15: ...p a minimum of 45 seconds for ISO accurate CO2 O2 N2O and agent readings 7 If not already set use the Display Setup Menu to select the gas waveforms to be displayed 8 If desired the gas waveform speed...

Page 16: ...accuracy using a full canister of Mindray DS approved precision calibration gas after calibration is performed Never use calibration gas that has expired has a different concentration or a canister th...

Page 17: ...appear At this point attach the calibration gas canister to the regulator and turn it on Increasing gas values will appear in the window as the Gas Module samples the calibration gas 6 When sampling i...

Page 18: ...bration Menu opens FIGURE 2 7 Calibration Menu 2 Select Gas Selection from the Calibration Menu and choose the calibration gas type Choices are Mixture 5 CO2 55 O2 33 N2O and 2 Des 3 Select Start to b...

Page 19: ...plete will be displayed next to each value that was successfully measured If at least one gas was successfully measured the Accept menu choice will become available If the values are acceptable select...

Page 20: ...stem detects a pneumatic leak Turn Gas Module and bedside monitor Off Also may appear when the Gas Module has been turned on without a sample line attached Install check sample lines filters water tra...

Page 21: ...cted or select Agent Auto ID GM Agent Mismatch ENF Appears when the system detects Enflurane as the primary agent and the manually selected agent is not Enflurane Match the Agent administered with the...

Page 22: ...act Mindray DS Technical Support GM Failed Appears when the Gas Module detects an unrecoverable error in its own operation Contact Mindray DS Technical Support GM Disconnected Appears when the bedside...

Page 23: ...ervice person only 2 The DRYLINE Water Trap Assembly consists of a filter housing and reservoir that must be checked and emptied whenever changing patients or if it is more than half full WARNING The...

Page 24: ...Operating Instructions NOTE Only the reservoir of the DRYLINE Water Trap Assembly may be cleaned and or disinfected NOTE If an Occlusion message appears it may be necessary to replace the DRYLINE Wat...

Page 25: ...0012 00 1276 01 Cable Gas Module to Bedside Monitor Serial Port long 1 8 m 0012 00 1276 02 Nasal Cannula CO2 7 2 1 m box of 10 0683 00 0424 10 Nasal Cannula CO2 O2 7 2 1 m box of 10 0683 00 0452 10 Ad...

Page 26: ...Accessories 4 2 0070 10 0696 02 Gas Module 3 Operating Instructions This page intentionally left blank...

Page 27: ...RH non condensing in Airway 0 100 RH condensing Operating Altitude Sea Level to 8 000 feet Shipping ISTA shipping procedure 1A Shock IEC 60068 2 27 peak acceleration 150 m s2 15 3 g duration 11 ms pul...

Page 28: ...ply with the following industry standards EN 60601 1 IEC 60601 1 UL 60601 1 CSA Standard C22 2 No 601 1M90 EN 60601 1 4 IEC 60601 1 4 ISO 21647 The Gas Module 3 has been certified by CSA Spillage and...

Page 29: ...r use in the electromagnetic environment specified below The customer or the user of the Gas Module 3 should assure it is used in such an environment EMISSIONS TEST COMPLIANCE ELECTROMAGNETIC ENVIRONM...

Page 30: ...al environment Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial or hospital environment Vo...

Page 31: ...engths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipme...

Page 32: ...IMUM OUTPUT POWER P OF TRANSMITTER IN WATTS W SEPARATION DISTANCE d IN METERS m ACCORDING TO FREQUENCY OF TRANSMITTER 150 kHz to 80 MHz d 1 2 x 80 MHz to 800 MHz d 1 2 x 800 MHz to 2 5 GHz d 2 3 x 0 0...

Page 33: ...t be liable for any incidental special or consequential loss damage or expense directly or indirectly arising from the use of its products liability under this warranty and the buyer s exclusive remed...

Page 34: ...appear to the Company to have been defective in material or workmanship No agent employee or representative of Mindray DS USA Inc has any authority to bind Mindray DS USA Inc to any affirmation repres...

Page 35: ...should be directed to the nearest Mindray DS location A list of international offices along with their phone numbers is provided at the end of this manual NOTE Upon request Mindray DS will provide ci...

Page 36: ...0070 10 0696 02 Rev C September 6 2010...

Page 37: ...Percy Road St John s Park Huntingdon Cambridgeshire PE29 6SZ United Kingdom Tel 01480 416840 Fax 01480 436588 Mindray Medical France SARL Europarc Cr teil 123 Chemin des Bassins 94035 Cr teil Cedex Fr...

Reviews: