background image

Gas Monitoring with Gas Module 3

Operations

2 - 2

0070-10-0696-02

Gas Module 3 Operating Instructions

WARNING: When using the Gas Module 3, the maximum sampling rate 

at the nasal cannula is 200 ml/min with an Adult/Pediatric 
water trap and 120 ml/min with a Neonatal water trap. 
This device should not be used on patients whose breathing 
could be impaired by this vacuum flow rate.

NOTE:

The Gas Module 3 is equipped with automatic barometric 
pressure compensation.

NOTE:

The Gas Module 3 uses a fixed correction of 11 hPa to 
compensate for the influence of water vapor in the gas 
sample, when converting the gas readings to ATPD. An 
increase in the ambient H

2

0 partial pressure to 30 hPa

(i.e. 28 °C, 80% RH or 33 °C, and 60% RH) will cause a 
general error for all gases of only -2% REL.

Monitoring Anesthetic Gases, O

2

, N

2

O and/or CO

2

NOTE:

To prevent moisture from entering the pneumatic system, 
ensure that the Gas Module 3 is always installed and 
operated in the horizontal orientation shown in all graphical 
depictions.

1.

Turn on the Gas Module 3 and the bedside monitor, and configure the bedside monitor 
serial port to be used with the Gas Module. Hold the 

DISCHARGE

 key in while 

powering unit on. Set alarms as desired.

FIGURE 2-2

Gas Module 3 Airway Adapter

NOTE:

DRYLINE

 Sample Lines are for use with Gas Module 3 only.

2.

For non-intubated patients, apply the nasal cannula to the patient. For intubated patients 
connect the sample line to the breathing circuit. Refer to instruction provided in the 
sample line packets.

3.

Connect the other end of the nasal cannula or sample line to the Gas Module at the 
input port. Ensure all tubing connections are tight.

Spectrum

S TANDB Y

DISCHARGE

D I S P L A Y

P R I N T

LE AD

S TA RT

TRENDS

FREEZE

INTERVAL

SIZE

S TO P

MA RK

EV ENT

STRIP

CO NT

E CG

P RINT

TREND

ECG

NIB P

A L A R M S

LIMITS

MUTE

MUTE

ALL

NORMAL
SCREEN

ZERO

ALL IBP

C.O.

PA WP

V IE W

NIBP: Idle

Parameters

Print Setup

Monitor Setup

Functions

Patient

HR

Source:Auto

E
C
G

S
T

N
I
B
P

C
O

T
e
m
p

bpm

0

mmHg

mmHg

mmHg

Sys/Dia

Sys/Dia

Source:CO2

Exp

Insp

0

/ 3 8

CPP

CPP

Mean

Mean

4

4

90

90

40

Resp

14 rpm

mmHg

-10

-10

10

10

mmHg

mmHg

121 81

Dia

Mean

Sys

/

mmHg

60

Mon

II

1 mV

Mon

I

1 mV

140

60

mmHg

S
P
O
2

C
O
2

I
C
P

C
V
P

P
A

A
R
T

180 120

Di a

MAP

Sys

CO

Interval: ON

ET:  13min

/

5.2

mmHg

l / min / m 2

CI

Lead

mm

II
I
V

+2.2
 - 2.1
 - 3.0

PVC/min: 45

JONES

, ERI

C

Adult

05/09/02

13.35

V
A S

122

170

55

%

SpO2

97

bpm

PR

60

Off

94

94

21

21

12

12

Dia

Dia

Mean

Mean

Sys

Sys

/

/

mmHg

mmHg

40

40

0

0

0

0

mmHg

mmHg

14

14

10

10

Sys/Dia

Sys/Dia

Mean

Mean

mmHg

mmHg

20

20

mmHg

mmHg

(14/16)

(14/16)

2.2

Source:Modul e

l / min

(5/2)

(5/2)

T1

37.0

T2

 T

36.9

0.1

C

7 Foot Nasal Cannula, CO

2

/O

2

P/N 0683-00-0452-10

7 Foot Nasal Cannula, CO

2

P/N 0683-00-0424-10

Adapter, Straight Tee ET
P/N 0683-00-0242-22

2.5 Meter Sample Line, Patient
Adult/Pediatric P/N 0683-00-0525-25
Neonate P/N 0683-00-0524-25

Adapter, Mask Elbow ET
P/N 0683-00-0242-12

Gas Scavenging Adapter Assy.
P/N 0997-00-0923
or
P/N 0997-00-0984

or

Summary of Contents for Accutorr 3

Page 1: ...Operating Instructions Gas Module To Purchase Visit Avobus com or call 1 800 674 3655...

Page 2: ...registered trademark of Mindray DS USA Inc Spectrum is a U S registered trademark of Mindray DS USA Inc Spectrum OR is a U S trademark of Mindray DS USA Inc Copyright Mindray DS USA Inc 2008 All righ...

Page 3: ...perations 2 1 Gas Monitoring with Gas Module 3 2 1 Pre use Test 2 3 Gas Monitor Calibration Passport 2 Spectrum Spectrum OR 2 4 Passport 2 Spectrum 2 4 Spectrum OR 2 6 Gas Module Troubleshooting 2 8 U...

Page 4: ...Gas Module 3 Operating Instructions 0070 10 0696 02 1 ii This page intentionally left blank To Purchase Visit Avobus com or call 1 800 674 3655...

Page 5: ...configuration of the bedside monitor Patents This device is covered under one or more of the following U S Patents 6 589 028 6 896 713 and foreign equivalents Possession or purchase of this device do...

Page 6: ...e Gas Module 3 on the floor Connect only a bedside monitor to the same MPSO as the Gas Module 3 Do not overload the MPSO WARNING Do not connect other equipment to the same MPSO with the Gas Module 3 a...

Page 7: ...damaged during operation there is a risk that bacteria and or mucus may contaminate the Gas Module WARNING Do not use Adult Pediatric type water traps and or sampling lines with neonates to avoid high...

Page 8: ...ca gel have been included CAUTION Contamination with CO2 N2O or Anesthetic Agent in the air surrounding the Gas Module 3 may cause significant measurement errors Notes NOTE Potential hazards due to er...

Page 9: ...rent AC Direct Current DC Data Input Output Interference may occur in the vicinity of equipment marked with this symbol Crossed out wheelie bin indicates separate treatment from general waste at end o...

Page 10: ...Gas Module 3 Operating Instructions 0070 10 0696 02 vii This page intentionally left blank To Purchase Visit Avobus com or call 1 800 674 3655...

Page 11: ...ir Adult Pediatric P N 0202 00 0182 10 Neonate P N 0202 00 0181 10 The Water Trap Assembly is used to capture moisture drawn in with the patient sample The Water Trap Reservoir must be emptied and rin...

Page 12: ...ower Cord 6 Exhaust Port This panel mount coupling is used for attaching a gas scavenging system P N 0997 00 0923 or P N 0997 00 0984 to the Gas Module 3 7 External Interface Port A communication inte...

Page 13: ...e monitors allows for the measurement of anesthetic gases O2 N2O and CO2 levels Measurement can be acquired via a nasal cannula non intubated for oxygen and CO2 only or through a sampling line connect...

Page 14: ...ient For intubated patients connect the sample line to the breathing circuit Refer to instruction provided in the sample line packets 3 Connect the other end of the nasal cannula or sample line to the...

Page 15: ...p a minimum of 45 seconds for ISO accurate CO2 O2 N2O and agent readings 7 If not already set use the Display Setup Menu to select the gas waveforms to be displayed 8 If desired the gas waveform speed...

Page 16: ...accuracy using a full canister of Mindray DS approved precision calibration gas after calibration is performed Never use calibration gas that has expired has a different concentration or a canister th...

Page 17: ...appear At this point attach the calibration gas canister to the regulator and turn it on Increasing gas values will appear in the window as the Gas Module samples the calibration gas 6 When sampling i...

Page 18: ...bration Menu opens FIGURE 2 7 Calibration Menu 2 Select Gas Selection from the Calibration Menu and choose the calibration gas type Choices are Mixture 5 CO2 55 O2 33 N2O and 2 Des 3 Select Start to b...

Page 19: ...plete will be displayed next to each value that was successfully measured If at least one gas was successfully measured the Accept menu choice will become available If the values are acceptable select...

Page 20: ...stem detects a pneumatic leak Turn Gas Module and bedside monitor Off Also may appear when the Gas Module has been turned on without a sample line attached Install check sample lines filters water tra...

Page 21: ...cted or select Agent Auto ID GM Agent Mismatch ENF Appears when the system detects Enflurane as the primary agent and the manually selected agent is not Enflurane Match the Agent administered with the...

Page 22: ...act Mindray DS Technical Support GM Failed Appears when the Gas Module detects an unrecoverable error in its own operation Contact Mindray DS Technical Support GM Disconnected Appears when the bedside...

Page 23: ...ervice person only 2 The DRYLINE Water Trap Assembly consists of a filter housing and reservoir that must be checked and emptied whenever changing patients or if it is more than half full WARNING The...

Page 24: ...Operating Instructions NOTE Only the reservoir of the DRYLINE Water Trap Assembly may be cleaned and or disinfected NOTE If an Occlusion message appears it may be necessary to replace the DRYLINE Wat...

Page 25: ...0012 00 1276 01 Cable Gas Module to Bedside Monitor Serial Port long 1 8 m 0012 00 1276 02 Nasal Cannula CO2 7 2 1 m box of 10 0683 00 0424 10 Nasal Cannula CO2 O2 7 2 1 m box of 10 0683 00 0452 10 Ad...

Page 26: ...Accessories 4 2 0070 10 0696 02 Gas Module 3 Operating Instructions This page intentionally left blank...

Page 27: ...RH non condensing in Airway 0 100 RH condensing Operating Altitude Sea Level to 8 000 feet Shipping ISTA shipping procedure 1A Shock IEC 60068 2 27 peak acceleration 150 m s2 15 3 g duration 11 ms pul...

Page 28: ...ply with the following industry standards EN 60601 1 IEC 60601 1 UL 60601 1 CSA Standard C22 2 No 601 1M90 EN 60601 1 4 IEC 60601 1 4 ISO 21647 The Gas Module 3 has been certified by CSA Spillage and...

Page 29: ...r use in the electromagnetic environment specified below The customer or the user of the Gas Module 3 should assure it is used in such an environment EMISSIONS TEST COMPLIANCE ELECTROMAGNETIC ENVIRONM...

Page 30: ...al environment Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial or hospital environment Vo...

Page 31: ...engths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipme...

Page 32: ...IMUM OUTPUT POWER P OF TRANSMITTER IN WATTS W SEPARATION DISTANCE d IN METERS m ACCORDING TO FREQUENCY OF TRANSMITTER 150 kHz to 80 MHz d 1 2 x 80 MHz to 800 MHz d 1 2 x 800 MHz to 2 5 GHz d 2 3 x 0 0...

Page 33: ...t be liable for any incidental special or consequential loss damage or expense directly or indirectly arising from the use of its products liability under this warranty and the buyer s exclusive remed...

Page 34: ...appear to the Company to have been defective in material or workmanship No agent employee or representative of Mindray DS USA Inc has any authority to bind Mindray DS USA Inc to any affirmation repres...

Page 35: ...should be directed to the nearest Mindray DS location A list of international offices along with their phone numbers is provided at the end of this manual NOTE Upon request Mindray DS will provide ci...

Page 36: ...0070 10 0696 02 Rev C September 6 2010...

Page 37: ...Percy Road St John s Park Huntingdon Cambridgeshire PE29 6SZ United Kingdom Tel 01480 416840 Fax 01480 436588 Mindray Medical France SARL Europarc Cr teil 123 Chemin des Bassins 94035 Cr teil Cedex Fr...

Reviews: