Gas Monitoring with Gas Module 3
Operations
2 - 10
Gas Module 3 Operating Instructions
GM: O
2
Uncalibrated
Appears after an unsuccessful
calibration attempt of the O
2
sensor.
The numeric data for O
2
will appear
as - - -, and the O
2
waveform will be a
flatline
Ensure proper gas mixture is
attached tightly and regulator
is on. Repeat calibration
procedure. If problem
persists, contact Mindray DS
Technical Support.
GM: N
2
O Uncalibrated
Appears after an unsuccessful
calibration attempt of the N
2
O
sensor. The numeric data for N
2
O will
appear as - - -, and the N
2
O
waveform will be a flatline
Ensure proper gas mixture is
attached tightly and regulator
is on. Repeat calibration
procedure. If problem
persists, contact Mindray DS
Technical Support.
GM: Agents Uncalibrated
Appears after an unsuccessful
calibration attempt of the agent
sensor. The numeric data for all
agents will appear as - - -, and the
agent waveform will be a flatline
Ensure proper gas mixture is
attached tightly and regulator
is on. Repeat calibration
procedure. If problem
persists, contact Mindray DS
Technical Support.
GM: Failed
Appears when the Gas Module
detects an unrecoverable error in its
own operation
Contact Mindray DS
Technical Support.
GM: Disconnected
Appears when the bedside monitor
cannot detect signals being sent by
the Gas Module
Ensure Gas Module is turned
on and interface cable is
properly connected. If
problem persists, contact
Mindray DS Technical
Support.
Sampling Error
Appears when a sampling error
occurs on one or more Gas Module
channels during calibration
Repeat calibration
procedure. If problem
persists, contact Mindray DS
Technical Support.
Not Ready For
Calibration
Appears when the Gas Module is
unable to initialize calibration
Repeat calibration
procedure. If problem
persists, contact Mindray DS
Technical Support.
Calibration Error,
Sampling Error
Appears when a sampling error
occurs in all four Gas Module
channels during calibration
Repeat calibration
procedure. If problem
persists, contact Mindray DS
Technical Support.
Calibration Error, Zeroing
Error
Appears when the Gas Module
cannot perform a
Zeroing
during
calibration
Repeat calibration
procedure. If problem
persists, contact Mindray DS
Technical Support.
MESSAGE/PROBLEM
REASON
SOLUTION