Gas Module 3 Operating Instructions
Cautions
CAUTION:
Use recommended Mindray DS supplied power cords. If a
substitute is necessary use only hospital grade power cords.
CAUTION:
The internal sampling system of the Gas Module does not
need to be cleaned or sterilized. There is no reverse flow
back to the patient. If the internal sampling system is
suspected to be clogged or dirty, the module should be
serviced by an authorized service person only.
CAUTION:
To avoid permanent damage, do not expose metal
components (pins, sockets, snaps) to disinfectants, soaps or
chemicals.
CAUTION:
Gas Module 3 must be moisture protected whenever
transported. This can be done with a protective plastic bag
in which water-absorbing materials (e.g. silica gel) have
been included.
CAUTION:
Contamination with CO
2
, N
2
O or Anesthetic Agent in the air
surrounding the Gas Module 3 may cause significant
measurement errors.
Notes
NOTE:
Potential hazards due to errors in software or hardware
have been minimized by actions taken in accordance with
IEC 60601-1-4.
Indication For Use
The indications for use for the Gas Module 3 include monitoring of airway gases during
anesthesia and/or assisted respiration. The intended environment of use is the anesthesia
department, including the Operating Room (OR) and post anesthesia care units (PACU), etc.
Unpacking
Remove the instrument from the shipping carton and examine it for signs of shipping
damage. Save all packing materials, invoice and bill of lading. These may be required to
process a claim with the carrier. Check all materials against the packing list. Contact the
Mindray DS Service Department at (800) 288-2121 or (201) 995-8237 for prompt
assistance in resolving shipping problems.