50
MW180 pH / mV / EC / TDS / NaCl / Temperature Bench Meter
12.1. TYPES OF LOGGING
Manual log on demand
y
Readings are logged each time LOG/CLEAR is pressed
y
All manual readings are stored in a single lot (i.e. records made on different days
share the same lot)
Log on stability
y
Readings are logged each time LOG/CLEAR is pressed and stability criteria is reached
y
Stability criteria can be set to fast, medium or accurate
y
All stability readings are stored in a single lot (i.e. records made on different days are
logged in the same lot)
Interval logging
y
Readings are logged continuously at a set time interval (e.g. every 5 or 10 minutes).
y
Records are added to it until the session stops.
y
For each interval logging session, a new lot is created.
A complete set of GLP information including date, time, range selection, temperature
reading and calibration information is stored with each log.
Manual Log on Demand
1. From the Setup mode, set Log Type to MANUAL.
2. From the measurement screen press LOG/CLEAR.
LCD displays “PLEASE WAIT”. The LOG ### “SAVED” screen displays stored log number.
“FREE” ### screen displays the number of available records.
Meter then returns to measurement screen.
Log on Stability
1. From the Setup mode, set Log Type to STABILITY and the desired stability criteria.
2. From the measurement screen press LOG/CLEAR.
LCD displays “PLEASE WAIT” then “WAITING”, until stability criteria is reached.
Note:
Pressing ESC or LOG/CLEAR with “WAITING” displayed, exits without logging.