MW180 pH / mV / EC / TDS / NaCl / Temperature Bench Meter
Overwriting existing data.
1. When the LCD displays “EXPORT” with ACCEPT and lot number blinking (USB icon
displayed), an identical named lot exists on the USB.
2. Press GLP/ACCEPT to continue. LCD displays “OVERWRITE” with ACCEPT tag blinking.
3. Press GLP/ACCEPT (again) to confirm. Not confirming exits the export.
Display returns to lot selection screen.
Data Management Warnings
“NO MANUAL / LOGS”
No manual records saved. Nothing to display.
“NO STABILITY / LOGS”
No stability records saved. Nothing to display.
"OVR" with lot ### (blinking) Identically named lots on USB. Select overwrite option.
"NO MEMSTICK"
USB drive is not detected. Data can not be transferred.
Insert or check the USB flash drive.
"BATTERY LOW" (blinking)
When low battery, export is not executed.
Recharge the battery.
Logged Data Warnings in CSV file
°
C ! Probe used beyond its operation specifications. Data not reliable.
°
C !! Meter in MTC mode.
°
C !!! Meter in NO TC mode. Temperature value only for reference.