MW180 pH / mV / EC / TDS / NaCl / Temperature Bench Meter
Individual Logs / Records
1. Press RCL to access the logged data.
LCD displays “PLEASE WAIT” followed by “LOG RECALL” with ACCEPT tag blinking and
the total number of logs.
2. Use
keys to select between log storage locations “PH” or “EC”.
3. Press GLP/ACCEPT to confirm.
4. Use
keys to select MANUAL or STABILITY lot type.
5. Press GLP/ACCEPT to confirm.
6. Use the
to navigate between logs. Log record number displays on the left.
7. With desired log record selected, press LOG/CLEAR to delete.
“DELETE” is displayed with ACCEPT tag and log ### blinking.
8. Press GLP/ACCEPT to confirm (to exit, press ESC or CAL/EDIT or LOG/CLEAR).
“DELETE” and Log ### blinking is displayed, until the log is deleted.
After the log has been deleted “CLEAR DONE” message displays briefly.
Display shows logged data of the next log ###.
Note:
Logs stored within an interval lot can not be deleted individually.
Log on Interval
1. Press RCL to access the logged data.
LCD displays “PLEASE WAIT” followed by “LOG RECALL” with ACCEPT tag blinking and
the total number of logs.
2. Use
keys to select between log storage locations “PH” or “EC”.
3. Press GLP/ACCEPT to confirm.
4. Use
keys to select an interval logging lot number.