52
MW180 pH / mV / EC / TDS / NaCl / Temperature Bench Meter
Interval Logging Warnings
“OUT OF SPEC”
Sensor failure is detected. Loggings stops.
“MAX LOTS”
Maximum number of lots reached (100). Cannot create new lots.
“LOG FULL”
Log space is full (1000 logs limit was reached). Loggings stops.
12.2. DATA MANAGEMENT
Each logging storage location (“PH” or “EC”) is independent and organized in lots.
y
A lot contains 1 to 600 log records (saved measurement data points)
y
Maximum number of lots that can be stored is 100, excluding Manual and Stability
y
Maximum number of log records that can be stored is 1000, across all lots
y
Manual and Stability logs can store up to 200 records (each)
y
Interval logging sessions (across all 100 lots) can store up to 1000 records.
When a logging session exceeds 600 records a new lot will be created.
y
Lot name is given by a number, from 001 up to 999. Names are allocated incrementally,
even after some lots have been deleted. Once lot name 999 was assigned, all lots have
to be deleted, to reset lot naming to 001.
See Deleting Data section.
12.2.1. Viewing data
1. Press RCL to access the logged data.
LCD displays “PLEASE WAIT” followed by “LOG RECALL” with ACCEPT tag blinking and
the number of stored logs.
2. Use
keys to select between log storage locations “PH” or “EC”.
Note:
Press RANGE
/
to export all logs from the selected location to external storage.
3. Press GLP/ACCEPT to confirm.
4. Use
keys to select the lot type (MANUAL, STABILITY or INTERVAL ###).
Note:
Press RANGE
/
to export only the selected lot to external storage.
5. Press GLP/ACCEPT to confirm.