3.  OPERATE WITH EasyCUT 

 
Place the microtome on the steel plate. 

 

 

Microtome not included in EasyCUT unit.  

 

Turn on EasyCUT, pressing the power button and wait for 4 seconds before starting the microtome cutting of 
the wax blocks. 

 

 
A flow of ionized air will start to come out from the four slots: 
 

 

 

 

Never  insert  objects  inside  the  slots  where  the  ionized  air  comes  out  from  otherwise 
electric shocks could occur. 

 
It is possible to turn  on the LED strip to light the cutting area. Press the  power button shown in  the  picture 
below:  

  

 

The distance between the blade of the microtome and the ionizer must not exceed 41cm if the 
unit is used at the minimum speed (1), otherwise it could not neutralize correctly. In this case it 
is recommended to increase the speed.

 

Summary of Contents for EasyCUT

Page 1: ...EasyCUT Static remover for microtome OperatorManual MM148 MILESTONE H E L P I N G P A T I E N T S ...

Page 2: ......

Page 3: ...ator manual and always keep it within reach for quick and convenient reference Please do not hesitate to contact our representative for any clarification or support you may require You can also contact us directly at Milestone s r l Via Fatebenefratelli 1 5 24010 Sorisole BG Italy Tel 39 035 412 8264 Fax 39 035 575498 Website www milestonemed com e mail marketing milestonemedsrl com Read carefully...

Page 4: ...2 2 Unpacking and check list 10 2 3 Placement 11 2 3 1 Unit overview 11 2 4 Power supply 13 3 OPERATE WITH EasyCUT 14 3 1 Adjusting the air flow direction 15 3 2 Adjusting the air flow speed 16 4 MAINTENANCE 18 4 1 Introduction 18 4 1 1 Monthly ionizing needle cleaning 18 4 1 2 Every 2 months fan filter cleaning 19 5 DECONTAMINATION AND CLEANING 21 5 1 Safety precautions 21 5 2 Procedure for EasyC...

Page 5: ... instruction comes with this symbol take great care as there is the risk of an electric shock Carefully follow the instruction in order to avoid this kind of accident Pay attention parts identified with this symbol are areas at risk of crushing hand and or fingers CE Logo this instrument complies with European Community directives It is the serial number Symbol of the European WEEE directive 2002 ...

Page 6: ...d comes out from that detect foreign objects in this case the inlet of ionized air is stopped The direction and speed of the ionized air can be easily adjusted at any time A series of LED shows the operating status at a glance Ionizing needles can be removed and installed and the air filter can be replaced very quickly Any other use of the instrument should be deemed as misuse and voids the manufa...

Page 7: ...r fuels could be present Never touch ionizing needles while the instrument is on electric shocks could occur please refer to chapter 2 3 1 Do not remove the instrument panels or cover Only authorized qualified and duly trained service personnel may repair the instrument and access the internal components It is important that normal standards for safety and good laboratory practices are applied Alw...

Page 8: ...pplication milestonemedsrl com Complying with directive 2012 19 EC of the European Parliament and of the Council of 4 July 2012 on waste electrical and electronic equipment WEEE the separate collection environmental managed of equipment is mandatory It is necessary to return the used equipment to the distributor or to inquire about the presence of a local empowered system for collection and dispos...

Page 9: ...g dimensions Width 560mm 22in Height 800mm 31in Depth 760mm 30in Unit dimensions Width 510mm 20in Height 670mm 26in Depth 590mm 23in Space dimensions required to place the unit Width 710mm 28in Height 680mm 27in Depth 600mm 24in Microtome available space Width 500mm 20in Height 578mm 23in Depth 469mm 18in The following pictures show the dimensions of EasyCUT and the space required UPPER VIEW The M...

Page 10: ...g The EasyCUT unit CODE SYN0001 includes the following items DESCRIPTION Qty IMAGE EasyCUT 1 Power cord 115V 1 Power cord 230V 1 Tool for speed adjustment 1 EasyCUT air grey filters for dust please refer to chapter 6 1 for spare parts codes 5 spare loose filters EasyCUT paper blue filters for air speed regulation please refer to chapter 6 1 for spare parts codes 10 spare loose filters EasyCUT MM14...

Page 11: ...UT in places where there are high levels of steam or dust in the air or where it could be exposed directly to water Do not place objects in the area under the ionized air outlet at a distance below 100 mm 4in see picture as they could jeopardize the proper operation and performances of EasyCUT 2 3 1 Unit overview The picture below shows the front view of the EasyCUT unit 1 Fan 2 Ionizers 3 LED lig...

Page 12: ...r comes out from or if the fan is obstructed 3 CHECK yellow light on shows that the inspection is required on the ionizing needles inserted in the 4 slots please refer to chapter A 1 or a decrease in the fan speed is occurring because of an obstruction or dirty filter please refer to chapter 4 1 2 for the filter cleaning 4 BAR ALARM it turns on if the error led or check led concerns the 4 slots wh...

Page 13: ...oard used in the laboratory must be compliant with the label positioned on the back of Milestone units Plug socket must be visible during normal use and easily accessible Before connecting the plug to the laboratory s socket check that the power supply voltage of the unit matches the power supply line of the laboratory Damage occurs if the instrument is supplied with incorrect power supply Power l...

Page 14: ...art to come out from the four slots Never insert objects inside the slots where the ionized air comes out from otherwise electric shocks could occur It is possible to turn on the LED strip to light the cutting area Press the power button shown in the picture below The distance between the blade of the microtome and the ionizer must not exceed 41cm if the unit is used at the minimum speed 1 otherwi...

Page 15: ... both hands turn the ionizing bar in the desired position paying attention not to insert fingers in the slots where needles are located Never insert objects inside the slots where the ionized air comes out from otherwise electric shocks could occur Use caution to avoid the risk of hand and or finger crushing during the ionizing bar movement 4 At the end keeping the ionizing bar in a hand move the ...

Page 16: ...tool for speed adjustment in the slot and turn clockwise as shown in the picture Minimum speed position 1 with blue filter Maximum speed position 4 without blue filter If the minimum speed position 1 with blue filter is not enough to neutralize the electrostatic charge remove the blue filter 1 Press in the top right corner of the fan as shown in the picture 2 Slide the fan grid to the left ...

Page 17: ...T MM148 001 Operator Manual 17 3 Remove the blue filter 4 Close the grid After that if the speed 1 without blue filter is not enough to neutralize the electrostatic charge increase the speed up to position 2 ...

Page 18: ... air comes out from otherwise the operation could be jeopardized When the product is used for long periods dust and other foreign particles can deposit on the ionizing needle and on the area around it If it happens clean the needles following the instructions below 1 Check that the power supply is disconnected 2 Remove the 4 blocks of ionizing needles from the back of the instrument as shown in th...

Page 19: ...the instrument before performing any procedure of assistance and maintenance When the product is used for long periods dust and other small foreign particles can deposit in the fan filter If it happens clean the air filter following the instructions below 1 Press in the top right corner of the fan as shown in the picture 2 Slide the fan grid to the left 3 Remove the blue filter ...

Page 20: ...with cleaning by removing the dust from inside the filter if the dust persists it is possible to clean the filter using water making sure that it is well dried before replacing it 5 It is now possible to insert the clean filter 6 Insert the NEW blue filter and close the grid ...

Page 21: ...y HI RISK single use protection gloves 100 latex or equivalent material Personal protection device compliant with the EN 374 standard and the 89 686 EEC Directive Antibacterial mask for surgical operating theatre use and white coat Personal protection device compliant with the EN 149 standard Grade FFP3D and the 89 686 EEC Directive In case of use of the personal protection equipment use products ...

Page 22: ...rine or its composed derivatives Never use abrasive sponges Do not lave steel in contact with steel wool or other metal objects that can contain or release oxidized particles Do not place objects exposed to high temperatures for long periods White covering support arm and housing of the ionizing bar Do not use solvents hydrocarbons and alcohol based detergents Never use abrasive sponges After clea...

Page 23: ...optional other than those provided by Milestone 6 1 Spare parts CODE DESCRIPTION QTY IMAGE SYN0023 EasyCUT air grey filters for dust qty 10 included 1 SYN0038 EasyCUT paper blue filters for air speed regulation qty 10 included 1 PN 50391 Power cord 115V 1 PN 50036 Power cord 230V 1 ...

Page 24: ... Check that the ionizer is properly installed If the ionizer is continuously subject to vibrations or collisions reassess the place of installation to reduce the quantity of vibrations and collisions Check that the switch to adjust the speed is set on a proper position not between two positions CHECK yellow led on BAR yellow led on Dirty ionizing needles Worn ionizing needles Turn off EasyCUT and ...

Page 25: ......

Page 26: ......

Page 27: ......

Page 28: ...marketing milestonemedsrl com www milestonemed com Via Fatebenefratelli 1 5 24010 Sorisole BG Italy Tel 39 035 4128264 Fax 39 035 575498 Milestone Srl MILESTONE H E L P I N G P A T I E N T S ...

Reviews: