2.4. Power supply
Power supply line needs to have a ground connection (Class I device).
Power line protection:
115V version: Miniature circuit breaker curve C, 15A, residual current device: 30mA.
230V version: Miniature circuit breaker curve C, 10A, residual current device: 30mA.
•
A ground connection is available in the lab to ensure the unit works correctly (yellow/green wire). Do
not connect a Milestone unit without ground connection. Do not use the blue/white wire (neutral of
power line).
•
Power supply line wires (their size and distance from the switch board) used in the laboratory must be
compliant with the label positioned on the back of Milestone units.
•
Plug/socket must be visible during normal use and easily accessible.
Before connecting the plug to
the laboratory’s socket, check that the power supply
voltage of the unit matches the power supply line of the laboratory.
Damage occurs if the instrument is supplied with incorrect power supply.
Power line protection:
230V: Differential circuit breaker curve C, 10A, breaking power greater than 6000A
Differential current: 30mA (Class I Device).
CONNECTING CABLES: Never connect or remove cables when power to the unit is
switched on, this will avoid risk of damage.
Connect the plug of the power cord to the EasyCUT socket and the plug of the main power cord to the main
power supply socket.