Intended use
9
This machine is programmed to carry out the final rinse with mains
water or with processed water of a quality to suit the application (e.g.
purified water, fully demineralised water or demineralised water).
It is particularly important to ensure the appropriate water quality for
the rinse and final rinse of items for analytical purposes.
The machine can be qualified for process validation.
The machine fulfills the requirements of Machinery Directive
2006/42/EC.
IMPORTANT
Australia and New Zealand
This machine is not intended to be used to reprocess or disinfect
medical devices or medical equipment.
Spray pressure and spray arm monitoring
The machine has a sensor for monitoring spray pressure in order, for
example, to detect pressure fluctuations due to misloading or foam in
the water circulation system. Spray pressure monitoring is set at the
factory to active in the "Cleaning" and "Final rinse" wash blocks.
The spray pressure monitoring result is documented together with
process documentation.
Spray arm speed can also be monitored, e.g. for detection of
blockages due to misloading or foam in the water circulation system.
Spray arm speed monitoring can be activated or deactivated via the
programmable settings.
Miele Service can make further settings for spray pressure and spray
arm monitoring.