Adding and dispensing chemical agents
51
Neutralising agent
Neutralising agent (pH setting: acidic) neutralises any residues of
alkaline cleaning agents on the surface of the load.
Neutralising agent is dispensed automatically in the
Interim rinse
phase after the main wash (see Programme charts). The reservoir
must be filled and the dispensing system vented for this to occur.
In the
Inorganic
programme neutralising agent is dispensed
additionally for an acidic pre-wash.
Refilling
neutralising agent
Place the neutralising agent container (red marking) on the open
cabinet door or on a surface which is robust and easy to clean.
Unscrew and remove the siphon. Place the siphon on the open
cabinet door.
Replace the empty container with a full one.
Push the siphon into the opening of the container and screw it back
on tightly. Observe the colour coding.
Wipe up any spilled chemical agent thoroughly.
Place the container on the floor next to the machine or in an
adjacent cupboard. The container must not be placed on top of or
above the machine. Make sure that the dispensing hose is not
kinked or trapped.
The dispensing system must then be vented (see "Settings
/DOS
venting").