2. SETTING OF NON-STANDARD INPUT

In parameter 

Sens

, equivalent to one of the non-standard inputs (name, starting with 

nS

),  the device  

uses polynomial calibration coefficients 'Non-standard input'. They convert normalized linear input 
0-10000

  (0-1V,   0-10V,   0-20mA   non-standard.),  μV(TC  non-standard),  Ω   (RTD400Ω   non-

standard), Ω/10 (RTD4KΩ non-standard),

 ==>

measured quantity

 by polynomial degree three. 

They   are   set   only   via   RS485   communication  with   device,   but   in   advance   must   be   calculated  
according to the characteristics of the sensor.

'non-standard' means sensor input with nonlinear transfer characteristic, for which is not pre-defined
in the device, i.e. it is absent from the values of parameter 

Sens

. This does not mean that the sensor

does not correspond to an industry standard. 

In sensor  TC  

non-standard

, the device measures the temperature of the cold end,  as for all other

thermocouples.

XIII.

RETURNING OF FACTORY SETTINGS

If calibration failed or incorrect configuration parameters it is possible to return the default settings,
by SYST, rE = 1 (hidden system parameters). 

Function only applies to certain parameters – hidden, service, calibration of input / output.

XIII.

MODBUS RTU RS485 COMMUNICATION (option)

The  device  is MODBUS RTU SLAVE  with possibility for communication  of 9600 or 19,200 bps  by

RS485 2WIRE line (parameter  

ConF

). In standard  performance  without using repeaters,  on  a line can  be

connected up to 32 devices, with repeaters — up to 247. There is an option for a special performance for up
to 128 devices without repeaters.

In the parameters table in column HOLDING REGISTER ADDRESS are the addresses of the relevant

parameters.  Here   will   be   considered   other   communication   options.  When   changing   the   parameter   or
operating mode there is an equality between change via the front panel buttons or change by network - ie the
device can be influenced by two sources simultaneously. If the network is a priority you can lock the keypad
(but it can be unlocked through the front panel).

IMPLEMENTED MODBUS FUNCTIONS

MODBUS

FUNCTION

COMMENT. RESTRICTIONS. 
Only specified with parameter and quantity names addresses can be used, and undescribed 
addresses in the manual should be considered as reserved addresses. The address fields in 
the table only indicate that the operation is feasible without EXEPTION (protocol error 
message).

01

Reading of single bits. 

03

Reading of HOLDING  REGISTERS,     0< REG ADR <127 – registers in the non-volatile 
memory,    128<REG ADR <256 – registers in RAM. When adding 512 - FLOAT 
calibration coefficients should be read in IEEE754 format (otherwise they are in format 
EXP, S.B0,B1,B2). 

05

Record of single bit,   439<COIL ADR<512;  967<COIL ADR<976

06

Record of single HOLDING REGISTER, 0<REG ADR<127 – in non-volatile memory; 

16

Record of many consecutive HOLDING REGISTERS. Action field as a function 06 plus 
addresses 512-640. In REG ADR>512 the function is to record the calibration coefficients 
FLOAT in IEEE754 format.

Document revision 2015-03-20

21

Summary of Contents for MS6026

Page 1: ...COMPETENCE IN MEASUREMENT Process Meter with Tricolor Bargraph and Universal Input MS6026 MS6126 v 2 2 USER MANUAL PLOVDIV 2015 Document revision 2015 03 20...

Page 2: ...wo limit values in the device 2 Setting of decimal point and lower and upper limit of the range set points parameters alarms bargraph etc 3 Selection of analogue output and range setting if it is avai...

Page 3: ...UT OFFSET 18 XI RECOMMENDATIONS AGAINST INTERFERENCE 19 XII ANALOG INPUTS SETTING SERVICE SETTING LINEAR INPUT SETTING BY TWO REFERENCE POINTS SETTING OF NON STANDARD INPUT 19 19 20 20 XIII RESTORING...

Page 4: ...Measurement accuracy 0 1 from range for 0 10V DC and 0 20 mA Time to refresh 0 5 sec Lower and upper limit NAMUR level detection 1 56 105 4 from the range i e for 4 20 mA 3 75 20 86 mA Auxiliary power...

Page 5: ...0 92 6 1950 24 0 15 0 04 Pt1000_385 EN 60751 200 850 185 2 3904 8 0 15 0 04 Pt100_391 GOST 200 850 17 24 395 16 0 15 0 04 Pt100_392 JIS C1604 81 200 660 17 08 337 03 0 15 0 04 Cu100_482 GOST 180 260 2...

Page 6: ...31 types of sensors 6 of which customer adjustable Universal power supply 90 250 VAC Tricolor bargraph 7 zones that can be configured by the customer as levels and color green red or orange Five digi...

Page 7: ...VI FRONT PANEL AND TERMINALS MS62026 Document revision 2015 03 20 7...

Page 8: ...N OF INPUTS AND OUTPUTS POWER SUPPLY Terminals 14 15 DIGITAL OUTPUTS Designation Model MS6026 MS6126 Contacts NO NO COM NC Outputs and Terminals 1 3 4 4 5 6 2 5 6 7 8 9 K3 7 8 10 11 12 K4 9 10 K5 11 1...

Page 9: ...input Positive terminal Iin PAS 25 Negative terminal Iin PAS 24 Voltage input 0 000 10 000 V Positive terminal U in 23 Negative terminal U in 24 Voltage input 0 000 1 000 V Positive terminal TC 29 Neg...

Page 10: ...t for MS6126 Alarm output K5 of MS6026 is identical to K3 of MS6126 Possible non standard versions in which K3 of MS8126 to realized as a control with hysteresis SP3 SP2 SP1 K3 K2 K1 3 3 1 1 SP4 K4 4...

Page 11: ...er Syst GC By SySt GP parameter the mode lighting column lighting point is determined When a LED is between two color zones when changing PV it would change its color thereby slightly expands one zone...

Page 12: ...rom parameters menu should not be in value change Press and hold this button for 4 sec while v 1_5 appears on the display V 1_5 is the software version of the device 2 minutes after the last pressed b...

Page 13: ...it is necessary points of AH and AL are automatically exchanged at rnGL rnGH 57 Uint 1 Hysteresis at 1 4 outputs operation A negative value inverts the corresponding output 1999 9999 Dimension as PV H...

Page 14: ...es if Syst Et 0 0 1V 0 10V 4 20mA and 0 20mA rnGL Measurement range lower limit Indication for 0V 0mA 4mA if Syst Et 0 1999 9999 Dimension as PV 20 Sint rnGH Measurement range upper limit Indication f...

Page 15: ...on linear input setting by two reference points If changed the current setting will remain valid until such new Et 1 or until range correction rnGL rnGH Et 0 BackUp options For Input Output settings a...

Page 16: ...tory set This setting scales the analog outputs to correspond to 0 20 mA and 0 10 V This requires an reference mA meter or V meter with accuracy higher than 0 1 to be connected to the respective outpu...

Page 17: ...s of the device 1 255 MODBUS HOLD REG ADDRESS 127 Uint 1 1 ConF System settings Field 1u indicates analog type used hardware related Field rr Normal Mode Rejection Ratio BAUDRATE PARITY STOP BIT are c...

Page 18: ...reviations Uint UNSIGNED INT Sint SIGNED INTEGER X OFFSET OF THE ANALOG INPUT CodE 47 In this mode can be entered value which will be added when measuring the input parameter offset if there is dispar...

Page 19: ...ly change of the input parameter but ignores short disturbances 3 Programming of Normal Mode Rejection Ratio NMRR filter Depending on the frequency of the power supply it is desirable to select approp...

Page 20: ...be entered in the lower range limit rnGL about 0 03 of the range For the range 1 V the optimum settings are made in range 0 1V non standard 1 SETTING OF LINEAR INPUT BY TWO REFERENCE POINTS CodE 80 T...

Page 21: ...be connected up to 32 devices with repeaters up to 247 There is an option for a special performance for up to 128 devices without repeaters In the parameters table in column HOLDING REGISTER ADDRESS a...

Page 22: ...LIBRATION COEFFICIENTS HOLDING REG ADR 512 524 626 628 TYPE FLOAT 4 BYTE IEEE754 RSP3 512 RSP2 514 RSP1 516 RSP0 518 Polynomial calibration coefficients Non standard input Convert normalized linear in...

Page 23: ...et 630 K_C4 K_C3 The device memory is RAM and non volatile Grey fields are part from RAM memory In power supply failure and restoring the power supply this quantities initialize equate their values fr...

Page 24: ...sport poor storage incorrect usage forces of nature failure to follow instructions and when others made an attempt to remove the defects In these cases the defect can only be removed for a fee Service...

Reviews: