CodE=81
]
[Clbr End]
Returning in the operating mode,
selecting of input Clbr End→ Clbr Intr → Pt 3 →10 v → 20 mA → Clbr End → ...
[Clbr Intr]
Internal calibration. Compensation points of 3-wire
RTD are shorted
(terminals 27,28).
After
confirmation [Clbr Uait] appears on the display. Wait
calibration to complete alone for ~ 10 s.
[Pt 3]
Reference resistance 3кΩ have to be connected,
4-wire.
[10 v]
Reference voltage 10V have to be connected
[20nA]
Reference current signal 20 mA have to be connected
Confirms the reference,
calibration is performed – [done]
appears on the display. Transition to
the next input.
Select the next input
without calibration
•
For the input range 10 V after setting is necessary to take account of the value in
the absence of input signal or with short on input terminals. Tthen the same value,
but with opposite sign must be entered in the lower range limit - rnGL / about
0.03% of the range.
•
For the range 1 V the optimum settings are made in range '0 -1V non - standard'.
1. SETTING OF LINEAR INPUT BY TWO REFERENCE POINTS – [CodE=80]
This setting destroys the service setting
Restoring of factory setting is possible using the parameter Syst, rE=1 from menu “Hidden system
parameters”
The type of setting is common for ranges
0-1V, 0-10V, 4-20mA
and
0-20mA
.
In order the function to be available, the parameter
Syst
,Et=1
,
The range rnGL,rnGH does not set readings, it is only warming.
This type of setting is appropriate when the parameters of connected transmitter are unknown and
could not be entered in
rnGL,rnGH
.
For access
Syst
,Et=1
, otherwise returns automatically in operating mode.
Enter
[CodE=80]
→
→ [tun1] appears on the display →
→ on input is fed first reference
value →
→ with buttons
is entered the fed value →
→ [tun2] appears on the
display →
→ second reference value is fed →
→ with buttons
is entered the fed
value →
→ operating mode.
Document revision 2015-03-20
20