Performing a Reference Analysis
Gemini VII 2390 Confirm
66 Gemini VII 2390 Confirm Installation Instructions and Checklist
Feb 09
2.
Place the analysis Dewar onto the elevator.
3.
Close the sample compartment door.
Starting the Reference Analysis
1.
Create a sample file for the reference analysis:
a. Select
New > Sample Information
; the Create a Sample File dialog is displayed.
b. Click the template named
Carbon Reference Material
, then
Create
. This template
contains the appropriate conditions for analyzing the carbon reference material; all
values will be copied into the file you are creating.