Gemini VII 2390 Confirm
Calibrating Transducers and Servo Valves
Feb 09 Gemini VII 2390 Confirm Installation Instructions and Checklist
51
f. Click
Next
. The following dialog is displayed.
g. When pressure stabilizes, click
From Reference
to enter the reading from the gauge,
or type in the pressure reading from the pressure calibration standard.
h. Press
Enter
.
i. Click
Close
when the calibration terminates.
j. Enable manual control and open valves 5-b and 5-s until the sample pressure is
between 0 and 300 mmHg. Close all valves.
k. Repeat step 2, but change the servo setting in step 2e to 380 mmHg.
l. Proceed if the values for the transducers and pressure gauge are within 2 mmHg of 380
mmHg. If the values are not within this range, repeat the calibration.