Performing a Blank Analysis
Gemini VII 2390 Confirm
60 Gemini VII 2390 Confirm Installation Instructions and Checklist
Feb 09
2.
Enter a description in the
Sample
field; for example, you may wish to enter the date. If
you include the dollar sign ($) as shown above, the automatically generated file number
will be included with the description.
3.
Click
Template
and double-click on the template named
Blank Tube Test
. This will
copy all of the appropriate values for performing a blank analysis into your sample file.
4.
Click
Report after analysis
and select
Report after analysis
. Choose
Preview
as the
destination, then click
OK
.
5.
Click
Start
; the User Log In dialog is displayed.
6.
Enter your password in the
Password
field and click
Ok
to begin the blank analysis.
7.
Review the analysis results to see if they are within the tolerances shown in the following
table. If these results are not achieved, perform a detailed leak check to determine the
cause.
P/Po
Tolerance in cm
3
0.1
±0.008
0.2
±0.010
0.3
±0.012
0.4
±0.014
0.5
±0.016
0.6
±0.018
0.7
±0.020
0.8
±0.022
0.9
±0.024