4
4. Remove close-fitting garments from the upper arm. To avoid
constriction, shirt sleeves should not be rolled up - they do not
interfere with the cuff if they are laid flat.
5. Always ensure that the correct cuff size is used (marking on the
cuff).
•
Fit the cuff closely, but not too tight.
•
Make sure that the cuff is positioned 2 cm above the elbow.
•
The
artery mark
on the cuff (ca. 3 cm long bar) must lie over
the artery which runs down the inner side of the arm.
•
Support your arm so it is relaxed.
•
Ensure that the cuff is at the same height as your heart.
6. Press the ON/OFF button
1
to start the measurement.
7. The cuff will now pump up automatically. Relax, do not move
and do not tense your arm muscles until the measurement
result is displayed. Breathe normally and do not talk.
8. When the correct pressure is reached, the pumping stops and the
pressure falls gradually. If the required pressure was not reached,
the device will automatically pump some more air into the cuff.
9. During the measurement, the pulse indicator
AO
flashes in the
display.
10.The result, comprising the systolic
AT
and the diastolic
AK
blood pressure and the pulse rate
AL
is displayed. Note also the
explanations on further display symbols in this booklet.
11.When the device has finished measuring, remove the cuff.
12.Switch off the device. (The monitor does switch off automatically
after approx. 1 min.).
You can stop the measurement at any time by pressing the
ON/OFF button (e.g. if you feel uneasy or an unpleasant
pressure sensation).
If the systolic blood pressure is known to be very high
,
it can be an advantage to set the pressure individually. Press
the ON/OFF button after the monitor has been pumped up to
a level of approx. 30 mmHg (shown on the display). Keep the
button pressed until the pressure is about 40 mmHg above
the expected systolic value – then release the button.
4. Appearance of the Pulse Arrhythmia (PAD)
This symbol
AR
indicates that certain pulse irregularities were
detected during the measurement. In this case, the result may
deviate from your normal blood pressure – repeat the measure-
ment. In most cases, this is no cause for concern. However, if the
symbol appears on a regular basis (e.g. several times a week with
measurements taken daily) we advise you to tell your doctor.
Please show your doctor the following explanation:
5. Traffic Light Indicator in the Display
The bars on the left-hand edge of the display
AM
show you the
range within which the indicated blood pressure value lies.
Depending on the height of the bar, the readout value is either
within the optimum (green), elevated (yellow), too high (orange) or
dangerously high (red) range. The classification corresponds to
the 4 ranges in the table as defined by the international guidelines
(ESH, ESC, JSH), as described in «Section 1.».
6. Data Memory
This device always stores the last result at the end of the measure-
ment. To recall the reading, press and hold the ON/OFF button
1
while the device is switched off. Release the button as soon as all
display elements are shown. The stored result will then be shown,
indicated with the letter «
M
»
BT
.
7. Battery Indicator and Battery change
Low battery
When the batteries are approximately ¾ empty the battery symbol
9
will flash as soon as the device is switched on (partly filled battery
displayed). Although the device will continue to measure reliably, you
should obtain replacement batteries.
Flat battery – replacement
When the batteries are flat, the battery symbol
9
will flash as soon
as the device is switched on (flat battery displayed). You cannot
take any further measurements and must replace the batteries.
1. Open the battery compartment
4
on the bottom of the device.
Information for the doctor on frequent appearance of the
Arrhythmia indicator
This device is an oscillometric blood pressure monitor that also
analyses pulse irregularity during measurement. The device is
clinically tested.
The arrhythmia symbol is displayed after the measurement, if
pulse irregularities occur during measurement. If the symbol
appears more frequently (e.g. several times per week on
measurements performed daily) we recommend the patient to
seek medical advice.
This device does not replace a cardiac examination, but serves
to detect pulse irregularities at an early stage.