Page 55
Serial Interface RS422
optoNCDT 1402
Error Output (Analog output)
8.3.11
Name:
SET_ANALOG_ERROR_HANDLER
Description:
Hold or not hold last measurement value.
Parameter:
X = 0 --> hold last measurement value
-
X = 1 --> error value (3.75 mA)
-
X = 2 … 99 --> hold last measurement value for 2 … 99 images respectively cycles
-
This command only affects the analog output. If set to X = 0, the last valid measurement value will continue
to be issued if an error occurs (no object, invalid object, object outside the measurement range or laser
turned off). If set to X = 1, an error signal will be generated for the current output that has an error value of
3.75 mA. If set to X = 2 ... 99, the last valid measurement value will continue to be issued for X measuring
cycles before an error signal is generated on the analog output.
Format: 31
24 23
16 15
8 7
0
hex
„+“
„+“
„+“
0x0d („CR“)
0x2B2B2B0D
„l“
„L“
„D“
„1“
0x494C4431
0x20
0x81
0x00
0x03
0x20810003
0x00
0x00
0x00
0x0X
0x0000000X
Reply:
31
24 23
16 15
8 7
0
hex
„l“
„L“
„D“
„1“
0x494C4431
0xA0
0x81
0x00
0x02
0xA0810002
0x20
0x20
0x0D
0x0A
0x20200D0A
Default setting:
error value