Page 33
Measurement Value Output
optoNCDT 1402
Calculation of a measurement value in mm from digital output:
Reference value Start of Measuring Range:
Reference value Midrange
x [mm] = (digital
OUT
* - 0.01) * MR [mm]
1.02
16368
x [mm] = (digital
OUT
* - 0.51) * MR [mm]
1.02
16368
Example: MR =10 mm, digital value = 2048, measurement value = 5 mm or 0 mm
Note: A digital value can be calculated from a measurement value (millimeter) as follows:
digital = + 0.01 *
OUT
x [mm]
MR [mm]
16368
1.02
Digital Error Codes
7.3
Digital error codes are issued in the same way as measurement values.
Value range for error codes: 16370 ... 16384 (digital
OUT
)
16370 no object detected
-
16372 too close to the sensor
-
16374 too far from the sensor
-
16376 target can not be evaluated
-
16380 target moves towards the sensor
-
16382 target moves away from sensor
-
16384 Error caused through image reading
-