Operation
capaNCDT 61xx/IP
6.
Operation
Connect the display/output devices through the signal output socket, see Chap.
. before connecting
the device to the power supply and switching on the power supply.
The measuring system is delivered calibrated. Calibration by the user is not necessary.
i
Allow the measuring system to warm up for about 10 minutes before the first measurement.
The power supply may not exceed or continuously fall below the specified limits.
Damage to or destruction of the sensor and/or controller.
Sensor
10 V
0 V
Target
Measuring range
Displacement, mm
100 %
0 %
0xF00000
0x0
RS485
1
20 mA
4 mA
1 = Start of
measuring range
2 = End of
measuring range
Fig. 22 Signal characteristic in the measuring range
Disconnect the power supply before touching the sensor surface.
> Static discharge
> Danger of injury
Digital interface with DT6120/IP/U or DT6120/IP/I only
Voltage or current output are only possible as an alternative.