Seite 14
Functional Principle, Technical Data
capaNCDT 61xx/IP
2.3
Technical Data
Controller modell
DT6110/IP/U
DT6110/IP/I
DT6120/IP/U
DT6120/IP/I
Resolution static
0.01 % FSO (2 Hz)
Resolution dynamic
0.02 % FSO (1000 Hz)
Frequency response
1 kHz (-3 dB)
Linearity
≤ ±0.1 % FSO
Sensitivity
≤ ±0.1 % FSO
Long-term stability
0.02 % FSO/month
Temperature stability
50 ppm FSO/K (+10 … +40 °C); 100 ppm FSO/K (-20 … +10 °C and +40 … +60 °C)
Operating
temperature
Sensor
type CSF -40 ... +100 °C | type CSE: -50 ... +200 °C
Sensor cable
permanent: -20 ... +80 °C, up to 10,000 h: -20 ... +100 °C
Controller
-20 … +60° C
Power supply
9 ... 28 VDC
(max. 1.4 W)
11 ... 28 VDC
(max. 2 W)
9 ... 28 VDC
(max. 1.6 W)
11 ... 28 VDC
(max. 2.2 W)
Output
analog
0 … 10 V
(short circuit proof)
optional: ±5 V, 10 … 0 V
4 … 20 mA
max. 500 Ohm load
0 … 10 V
(short circuit proof)
optional: ±5 V, 10 … 0 V
4 … 20 mA
max. 500 Ohm load
digital
---
---
RS485, 230400 Baud (adjustable),
24 bit measuring values, max. 2 kSamples (adjustable)
Storage temperature
-20 ... +75 °C
Protection
class
Sensor
IP 40, IP 54 (functionality reduced)
Controller
IP 68
Shock
20 g / half sine 5 ms in XYZ axis / 1000 shocks per axis (DIN-EN 60068-2-29)
Vibration
10 g / 10 … 2000 Hz in XYZ axis / 10 cycles per axis (DIN-EN 60068-2-6)
Weight
165 g
FSO = Full Scale Output