Acquiring the Dark Signal

boreCONTROL

5. 

Acquiring the Dark Signal

The dark signal of the sensor represents an intrinsic offset level generated by parasitic 
reflections inside the controller, which must be taken into account for the controller to 
be able to operate correctly. The level of the dark signal depends on the sampling rate. 
The dark signal should be acquired at all sampling rates in order to be able to subtract 
it while the controller is measuring. A dark signal acquisition is performed during adjust-
ment by the manufacturer, but must be repeated at regular intervals.

  If you change the light intensity at the external light source or you change the sen-

sor, you have to repeat the dark signal measuring.

i

  After each sensor change the dark signal has to be measured for the first time.

The dark signal compensation is described in the software manual.

Summary of Contents for boreCONTROL

Page 1: ...Measuring system for internal wall inspection in bore holes Instruction manual boreCONTROL...

Page 2: ...O EPSILON MESSTECHNIK GmbH Co KG K nigbacher Stra e 15 D 94496 Ortenburg Tel 49 8542 168 0 Fax 49 8542 168 90 email info micro epsilon com www micro epsilon com Certified according to DIN EN ISO 9001...

Page 3: ...n Unit BCM2410 8 4 2 Controller BCC2410 9 4 3 IFC24x1 10 4 4 Electrical Connections BCC2410 10 4 4 1 Power Supply 10 4 4 2 Rotation Unit BCM2410 11 4 4 3 IFC 24x1 11 4 4 4 RS232 11 4 5 Electrical Conn...

Page 4: ...boreCONTROL...

Page 5: ...ry Damage to or destruction of the controller Avoid shock and vibration to the sensor and controllers Damage to or destruction of the sensor and controllers Avoid exposure to aggressive materials wash...

Page 6: ...158 F Humidity 5 95 non condensing Ambient pressure atmospheric pressure EMC According to EN 61000 6 3 2007 EN 61000 6 2 2005 Functional Principle Technical Data 2 Short Description 2 1 The rotating...

Page 7: ...sity max Intensity min Intensity ok Out of range Midrange In range Power On External light source Status In Ethernet External light source Sensor RS422 Encoder Dark reference STATUS PS2020 230 VAC PE...

Page 8: ...e contact the manufacturer or supplier Storage 3 2 Storage temperature 0 C up to 75 C Relative humidity 0 95 non condensing Mounting 4 The sensors of the series BCS241x x are optical sensors which are...

Page 9: ...rawing mounting adapter dimensions in mm inches not to scale Controller BCC2410 4 2 Keep the connectors free in mounted position Mount the controller on a DIN rail type TS35 R5 84 3 31 76 15 3 00 83 7...

Page 10: ...r Supply 4 4 1 The plug in terminal strip is designed for a conductor cross section of 0 25 mm to 1 0 mm The terminal strip is mounted on the controller with two screws and can be re moved for wiring...

Page 11: ...ensity ok Out of range Midrange In range Power On External light source Status In Ethernet External light source Sensor RS422 Encoder Dark reference STATUS IFC Sensor controller connected to the boreC...

Page 12: ...ernet External light source Sensor RS422 Encoder Dark reference STATUS PS2020 230 VAC PE N L Power supply of the boreCONTROL with a PS2020 Fig 9 Ethernet 4 5 2 Connect the controller connector Etherne...

Page 13: ...ror 1 GND Shield Error 2 GND HLL 24 VDC GND Shield U out GND U Shield I out GND I Power Digital I O RS422 Encoder Analog Out confocalDT Intensity max Intensity min Intensity ok Out of range Midrange I...

Page 14: ...from the packing Move the sensor into the rotation unit Turn the lock ring 90 degrees to the left Remove the disassembly aid in hori zontal direction Rotate the sensor until the marks on the sensor an...

Page 15: ...dismounting is described Slide the disassembly aid in hori zontal direction on the sensor Lock ring Sensor confocal Disassembly aid Turn the lock ring 90 degrees to the right Now the sensor is locked...

Page 16: ...g rate The dark signal should be acquired at all sampling rates in order to be able to subtract it while the controller is measuring A dark signal acquisition is performed during adjust ment by the ma...

Page 17: ...damages MICRO EPSILON always strives to supply the customers with the finest and most advanced equipment Development and refinement is therefore per formed continuously and the right to design changes...

Page 18: ...EMC requirements Part 1 General requirements We also declare that the special technical documentation for this partly completed ma chine has been created in accordance with Annex VII part B and commi...

Page 19: ......

Page 20: ...MICRO EPSILON MESSTECHNIK GmbH Co KG K nigbacher Str 15 94496 Ortenburg Germany Tel 49 0 8542 168 0 Fax 49 0 8542 168 90 info micro epsilon de www micro epsilon de X9751231 B02 X9751231 B021012HDR...

Reviews: