Page 11
Mounting
boreCONTROL
Rotation Unit BCM2410
4.4.2
Connect the controller (connector “Sensor“) with the rotation unit. Please use the
SC2410-03 cable.
+24V DC GND
RS232
IFC
Sensor
bore
CONTROL
Rotation unit connect to the boreCONTROL
Fig. 7
IFC 24x1
4.4.3
Connect the controller (connector “IFC“) with the IFC24x1. Use the SC2410-0,5
cable.
bore
CONTROL
Sensor
+24V DC GND
RS232
SyncIn / T
rigIn
GN
D
Shield
SyncOut
GN
D
Er
ror 1
GN
D
Shield
Er
ror 2
GN
D
HLL
24 VDC
GN
D
Shield
U out
GN
D U
Shield
I out
GN
D I
Power
Digital I/O
RS422
Encoder
Analog Out
confocal
DT
Intensity >max
Intensity <min
Intensity ok
Out of range
Midrange
In range
Power On
External light
source
Status In
Ethernet
External
light source
Sensor
RS422
Encoder
Dark reference
STATUS
IFC
Sensor controller connected to the boreCONTROL
Fig. 8
RS232
4.4.4
Connect the controller (connector
“
RS232“) with the PC. Use a standard interface
cable.
PIN No.
9-pin Sub-D
Assignment
+24V DC GND
RS232
IFC
Sensor
bore
CONTROL
USB 3/4
USB 5/6
Digital I/O
Analog I/O
Firewire / IEEE139
4
1 2 3
USB 1/2
COM1
Power
IPC-4
VGA
LAN1
LAN2
DVI
Relais
Remote
2
TxD
3
RxD
5
GND
Via the RS232 interface, the instruc-
tions of the motor control are trans-
ferred to the rotation unit.
Baud rate: 9.6 kBaud. Data format:
8 data bits, no parity, one stop bit
(8,N,1).
The reference grounds (GND) of con-
troller and receiver (e.g. PC or PLC)
must be connected.