Page 9
Functional Principle, Technical Data
optoNCDT ILR2250
3.4
Technical Data
Model
ILR2250-100
ILR2250-100-IO
Item number
7112015
7112016
Measuring range
1
SMR
EMR
SMR
EMR
black 6 %
0.05 m
30 m
0.05 m
30 m
gray 40 %
0.05 m
70 m
0.05 m
70 m
white 80 %
0.05 m
100 m
0.05 m
100 m
Reflector film
2
35 m
150 m
35 m
150 m
Measuring rate
20 Hz
Resolution
0.1 mm
Linearity
< ±1 mm
3
Repeatability
4
< 300 µm
Temperature compensation
-10 ... +50 °C
Light source
Semiconductor laser < 1 mW, 655 nm (red)
Typ. service life
50,000 h
Laser safety class
Class 2 according to DIN EN 60825-1: 2015-07
Permissible ambient light
50,000 lx
Supply voltage
10 … 30 VDC
Power consumption
< 1.5 W (24 V)
Signal input
Trigger
-
Digital interface
RS422 / USB
5
IO-Link 1.1; process data,
parameter set up and diagnostics
Analog output
4 ... 20 mA (16 bit, freely scalable with-
in the measuring range)
-
Switching output
Q1 / Q2 / Q3 (configurable)
Q1 / Q2 / Q3 (configurable) included in
IO-Link process data
Connection
Supply/signal: 12-pin M16 screw/plug
connection
(see accessories for connection cable)
Supply/signal: 5-pin M12 screw/plug
connection
(see accessories for connection cable)
Mounting
Screwing and adjustment on sensor base plate
Temperature range
Storage
-25 ... +70 °C (non-condensing)
Operation
-10 ... +50 °C (non-condensing)
Shock (DIN EN 60068-2-29)
15 g / 6 ms in 3 axes, in 3 directions, 1000 shocks each
Vibration (DIN EN 60068-2-6)
15 g / 10 … 500 Hz in 3 axes, 10 cycles each
Protection class (DIN EN 60529)
IP65
Material
Aluminum housing
Weight
approx. 300 g
Control and display elements
5x LEDs for power, signal strength and switching outputs
Features
4 measurement-specific operating
modes via sensorTOOL
4 measurement-specific operating
modes via IO-Link
SMR = Start of measuring range, EMR = End of measuring range
The specified data apply for a consistent room temperature of 20 °C, sensor is continuously in operation. Measured on
white, diffuse reflecting surface (reference ceramic)
1) Depends on the reflectivity of the target, ambient light interference and atmospheric conditions
2) ILR-RF210 reflector film 210 x 297 mm; art. 7966058
3) Measured in the range of 0.05 … 20 m; statistical spread 2s
4) Measurement frequency of 20 Hz, moving average 10
5) Connection via interface module, see accessories