5.11 Life-Stat
®
Model 1008
MII
Detailed Specifications
Input:
- Compressed O
2
at 50 to 90 psi (3.515 to 6.327 kgf/cm
2
)
- Gas Consumption (average)*:
in 30:2 mode- 34 L/min (8.98 gal/min) @100 CPM**, 70 L/min (11.62 gal/min) @120 CPM***
in CCV mode- 44 L/min (11.62 gal/min) @100 CPM**, 134 L/min (35.4 gal/min) @120 CPM***
- Indicator to show adequate input pressure: 50
3 psi (3.515
0.211 kgf/cm
2
)
- Pressure relief valve set at 100
5 psi (7.030
0.351 kgf/cm
2
)
- Filters to prevent contamination
- Oxygen checked quick connector (provided on O
2
Input Hose)
Compression:
- Compression Frequency: 100 or 120
2 compressions per minute***
- Compression Stroke Range: Continuously Adjustable, 0 to 8 cm (0.0 to 3.15 in)
- Compression to Ventilation Ratio: 30:2 (30:2 mode)
- Continuous Compression with simultaneous Ventilation (CCV mode)
- Relaxation Force range: Upstroke force of at least 1.361 kg (3.0 lbs)
- Duty Cycle: constant at 50:50 (Systolic:Diastolic)
- Chest Compression waveform:
Exponential waveform with a time constant of less than 60.0 msec
Ventilation:
- Time-cycled, constant-flow ventilator
- Calibrated Volume Control: 0-1000
100 mL; adjustable from 0 - 1000 mL
(0. - 0.317
0.026 gal; adjustable 0 - 0.317 gal)
- Deliver 95%
5% Oxygen to the patient
- Inspiratory Time: 1.0
0.25 seconds
- Inspiratory Flowrate: 0 to 50.0
2.5 LPM (0 to 13.20
0.66 gal)
- Inspiratory:Expiratory (I:E) ratio: (30:2) – 1:1.5 @ 100 compressions per minute***
- Inspiratory:Expiratory (I:E) ratio: (CCV) – 1:5.5@ 100 compressions per minute***
- Inspiratory:Expiratory (I:E) ratio: (30:2) – 1:1.3 @ 120 compressions per minute***
- Inspiratory:Expiratory (I:E) ratio: (CCV) – 1:4.6 @ 120 compressions per minute***
- Pressure Relief Valve: < 55.0 cm H
2
0 (0.782 psi)
Controls:
- System Control membrane switch pad
START/ON / 30:2 / CCV / PAUSE / STOP/OFF
- Compression Control: Continuous Compression Depth
- Ventilation Control: Calibrated Tidal Volume
- Low Battery indicator
- Service interval indicator
Environmental:
- Operating Environment: -20
C to 45
C (-4
F to 131
F)
- Storage Environment: -30
C to 60
C (-22
F to 140
F)
- Humidity: 0 to 98% RH (non-condensing)
- Sealed piston shaft and bearing to prevent contamination
Dimensions:
Height: 22-1/2” (57.2cm) Width: 7-5/8” (19.4cm) Length: 18-1/4” (46.4cm)
Weight: 16 lbs (7.26kg) 19.5 lbs (8.85kg) (with base attached)
Questions about the Life-Stat
®
? Please call 1-800-530-9939.
*Based on compression depth of 5cm at 500mL ventilation volume. Actual consumption varies depending on device settings.
**CPM (compressions per minute)
*** Depending on version ordered. Devices set at 120 CPM will have a serial number of LS-4000 or higher.