Factory-Recommended Maintenance/Service Intervals
Life-Stat
®
Use
Shift Check
Functional Check
Factory
Service
Heavy use: Daily
After each use
Monthly
Every 2
years
Frequent use: 10 - 15 times per month
After each use
Quarterly
Every 3 - 4
years
Infrequent use: < 10 times per month
After each use
Semi annually
Every 5
years
(Use based on a 20 minute average run time.)