REV: 2019-04
20
3.4
Life-Stat
®
Deployment
3.4.1 Life-Stat
®
Setup:
A.
Remove the Life-Stat
®
from the Case.
B. Ensure the System
C. Ensure Controls (#2) &
Control (#1) is OFF. (#3) are both turned fully
counterclockwise.
D. Attach the Breathing Hose/Non Re-breathing valve (and mask,
if used) to the PDV.
E.
Ensure the Oxygen source is energized then attach the O
2
Supply Hose to the Oxygen source. Pull the collar back from
the O
2
Source-end connector, press firmly onto the male
connector of the Oxygen source, then release the collar to
secure the connector. Pull gently on the hose to ensure a secure
connection.
F.
Attach the opposite end of the hose to the Life-Stat
®
O
2
Supply Connector by inserting the hose
connector while slightly turning it at the same time
to align the hexagons, and then press firmly to attach. Pull
gently on hose to ensure a secure connection. Do not press the
release button while attaching the connector. Only press the
release button to disconnect the connector.
G.
Verify that the green Pressure Indicator located on the top of the
column shows an adequate input pressure is available.
Press for
OFF