2.3.1 Life-Stat
®
System
The Arm and Column positions the Piston and Massager Pad correctly over the patient’s sternum. It is
designed to provide a sternal deflection percentage based on the patient A-P chest diameter. Sternal
deflection is nominally set to 20% of the A-P diameter. The depth of each chest compression is easily
monitored using the markings on the Dome surrounding the Piston. The Column also serves as a
storage tank that holds sufficient Oxygen to continue to drive the Life-Stat
®
for several compressions
during an Oxygen source change.
Piston
(inside Dome)
Massager
Pad
Handle
Arm
Lock
Dome
Sternal
Deflection
Number
A-P Diameter
Sternal Deflection
Number
Arm
Column
O
2
Supply
Hose
Patient
Demand Valve
(PDV)
O
2
Supply Connector
Battery
Compartment
(1 of 2)
Base
The Life-Stat
®
System
Pressure
Indicator
BackBoard