REV: 2019-04
12
2.
COMPRESSION DEPTH (Control #2):
This control is used
for setting the depth of compression on the patient. The depth is
set to correspond to the measured A-P diameter shown on the
scale located on the back of the Column. As indicated by the
arrow, turn the knob clockwise to increase the compression
depth and counterclockwise to decrease compression depth.
NOTE:
After the massager pad has been lowered to the patient's chest,
the correct patient A-P
chest diameter is determined by locating the number on the back of the Column (A) just above
the Arm where the white arrows (B) are located. Set the compression depth to the
corresponding indicator number on the Dome (C) to match the A-P diameter number
indicated on the Column.
3.
VENTILATION VOLUME (Control #3):
This control
adjusts the tidal volume (0-1000mL) of Oxygen delivered to
the patient from the PDV. The tidal volume is increased with
a clockwise rotation and decreased with a counterclockwise
rotation. With the control in the fully counterclockwise
position, no (0mL) tidal volume is delivered.
WARNING: Before operation, ensure that all controls are in the “OFF” or
“decreased” (fully counterclockwise) position before connecting to an Oxygen supply
or placing the unit on the patient. By verifying the position of all controls the
operator is assured of proper operation.